Vötsch VCE 1000 Classic Emission Testing Chamber FID 2000MP


Item ID: 4338

Condition: Used

Stock: Only ${ $store.getters['4338/currentItemVariation'].stock.net } in stock

The item is already sold

* Excl. VAT excl. Shipping

Vötsch VCE 1000 Classic Emission Testing Chamber with Microprocessor-Controlled Total Hydrocarbon Analyzer FID 2000MP


Here we offer a Vötsch VCE 1000 Classic Emission Testing Chamber + Microprocessor-Controlled Total Hydrocarbon Analyzer FID 2000MP
VCE Emission Testing Chamber VCE 1000 with enhanced performance
and separate control cabinet
Designation: VCE 1000 Classic
Test chamber volume (usable) l 1000 (916)
Performance for temperature tests:
Temperature range °C +20 to +130
Temperature deviation over time K ±0.1 to ±0.3
Temperature deviation spatially K ±0.5
Temperature gradient K 1
Temperature change rate:
Cooling K/min 0.3
Heating K/min 0.4
Performance for climate tests
Conditioning:
Carrier gas conditioning with climate module
according to the evaporation principle
Temperature range °C +20 to +130
Humidity range % 5 to 95
Dew point range °C +5 to +60
Humidity deviation over time % ±1 to ±3
Carrier gas exchange rate 0.1 to 1.8 chamber volumes per hour
Intake condition min. +10 °C / 50 % r. H.
max. +32 °C / 80 % r. H.
Performance range for desorption
Desorption temperature up to max. +240 °C, adjustable
Test chamber surface temperature between +220 °C and +240 °C
Purge process max. 4 m³/h
Test chamber dimensions Width mm 750
Depth mm 1630
Height mm 750
Test chamber door mm 670x670
External dimensions Width mm 1350
Depth mm 2950
Height mm 2050
Sound pressure level 2) dB(A) 63
Nominal power kW 10.0
Electrical connection 3/N/PE AC 400 V ± 10 %, 50 Hz
+
FID 2000 MP
Microprocessor-Controlled Total Hydrocarbon Analyzer
Product description:
The Flame Ionization Detector 2000MP measures the
total hydrocarbons in SHED chambers
of various sizes.
Special advantages:
• Optimal sampling volumes
• Up to 4 test gases can be connected
• Heating time from 20°C to 200°C in just
30 minutes
• Automatic calibration
• Automatic flame ignition
• Calibration without measurement value change
Technical data FID 2000MP
19" Rack version
Measuring component: CxHy
Measurement display: 3 1/2 digit, LED
Measurement ranges: 4
Smallest measurement range: 0 - 10 ppm
Largest measurement range: 0 – 3,000 ppm
Range switching: Manual / Computer
Reproducibility: +/- 1 %
Zero point drift: +/- 1 % in 24 hours
Response time 1 sec. (Minished)
from device input:
Heating time approx. 30 min
Analog outputs:
- Current, galvanically isolated: 0-20 mA, 4-20 mA
- Voltage: 0-10 V
Status message: Individual messages
Negative pressure FID: 0.4 bar vacuum
Auxiliary gases:
- Fuel gas He/H2, 5.0
- Test gas: C3H8
- Zero gas: N2, 5.0 or
synthetic air
- Combustion air: Combustion air generation
Fuel gas consumption: 150 ml/min
Combustion air consumption: 500 ml/min
Zero and test gas consumption: 1000ml/min
Power supply: 230 V / 50 Hz
Power consumption: 300 W
Ambient temperature: 0 - 45°C
Dimensions (H x W x D): 3HE x 19"x460 mm
Weight: approx. 25 kg
Scope of delivery. (See image)

The device comes from a stock liquidation.

We deliver with express shipping (DHL, UPS, TNT)
 
Further questions can be answered by phone.

Written orders possible via email or fax

Technical characteristic Value
Item ID 4338
Condition Used
Model VCE 1000
Manufacturer Vötsch
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[4338].variation.documents[0].data.item.id;

EU Responsible Person

Weiss Technik GmbH
Greizer Straße 41-49
35447 Reiskirchen Germany
info@weiss-technik.com
+49 6408 84-0

Manufacturer details

Vötsch
Weiss Technik GmbH
Germany

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system

Climate test cabinets