Type: WK11-600/70/5
Weiss climate test chamber water-cooled
Simcon/32 – Net controller
Technical data:
Test chamber designation: WK11-600/70/5
External dimensions including hinges and door locking system:
Height 1995 mm, Width 1090 mm, Depth 1660 mm
External dimensions excluding hinges and door locking system:
Height 1995 mm, Width 1000 mm, Depth 800 mm
Internal dimensions:
Height 950 mm, Width 800 mm, Depth 800 mm
Feed-throughs:
2 × R1 DN 125 mm Ø
1 × L1 DN 50 mm Ø
Test chamber volume: approx. 600 liters
Data for temperature tests:
Temperature range: −70 °C to +180 °C
Temporal temperature deviation (center of usable space): ±0.1 to ±0.5 K
Spatial temperature deviation: ±0.5 to ±2 K
Temperature gradient: 1 to 4 K
Temperature change rate:
Heating: 6 K/min
Cooling: 6 K/min
Heat compensation:
at +20 °C: 5000 W
at −20 °C: 5000 W
Data for climate tests:
Temperature range: +10 °C to +95 °C
Humidity range: 10 % to 98 % RH
Temporal temperature deviation (center of usable space): ±0.1 to ±0.3 K
Spatial temperature deviation: ±0.5 to ±1 K
Temperature gradient: 1 to 2 K
Temporal humidity fluctuation (center of usable space): ±1 to ±3 % RH
Heat compensation: 500 W
3/N/PE AC 400 V ±10 % 50 Hz
Rated power: 16.5 kW
Rated current: 29 A
On-site fuse protection: 32 A slow-blow
Permissible test specimen weight:
Total load on shelf supports: 80 kg
Test chamber floor: 80 kg, insert shelf: 40 kg
Weight: 675 kg
Refrigerant: R452A / R23
Important notice:
“The table shown in the photos is not included in the sale.”
For your safety as a buyer, please note the following information:
The following procedures are carried out on all climate chambers offered by us prior to sale:
1. Functional inspection and replacement of required components
2. Recharging with legally compliant refrigerant if necessary
3. Leak test including certificate
4. Documented test run after successful inspection
Condition: used
Scope of delivery: (see photos)
(Subject to changes and errors in technical data and specifications!)
If you have any further questions, please feel free to contact us by phone.
Switch to environmentally friendly refrigerant
The following work can be performed on climate chambers. Please contact us if you are interested.
- Extraction and proper disposal of the previous refrigerant
- Replacement of necessary components, if unavoidable
- Evacuate the circuit and refill it with R 452A, for example.
- Checking / adjusting overheating in expansion valves
- Leak test
- Test run of the system