Test cabinet for lithium-ion energy storage BINDER KB 720 with equipment package P Plus


Item ID: 24948

Condition: Used

Stock: Only ${ $store.getters['24948/currentItemVariation'].stock.net } in stock

EUR 14,200.00 *
Content 1 piece
Unit price EUR 14,200.00 / piece
The item is in stock

PAYMENT METHODS:

²


* Excl. VAT excl. Shipping

Binder


BINDER KB 720 Cooling Incubator with Package P Plus
– Battery Test Cabinet EUCAR 5, CO₂ Fire Suppression System
Device Description:
For sale is a BINDER KB 720 (E6) cooling incubator, manufactured in 2022,
equipped with the enhanced safety and performance test package "P Plus".
This device is not a standard laboratory incubator, but a high-quality safety test cabinet,
specifically designed for aging, performance, and safety tests of lithium-ion energy storage systems.
Special Value:
The device is designed to safely detect,
monitor, and contain critical battery events
– up to EUCAR Hazard Level 5 (fire/flames, no explosion).
What can the device do specifically? (Point 2 – Performance & Application)
Conduct aging tests on lithium-ion cells
Performance tests (charge/discharge cycles under temperature)
Temperature endurance runs and temperature cycling tests
Tests of unpowered and cyclically powered cells
Reproducible, precise temperature control throughout the usable space
Safe execution of tests under realistic thermal conditions
Test environment for R&D, automotive, e-mobility, and energy storage technology
Suitable for:
Battery & cell manufacturers
Research and development laboratories
Testing laboratories & safety testing
Automotive / E-Mobility / Energy Storage

The P Plus package turns the KB 720 into a true battery safety testing cabinet:
Integrated safety features:
Automatic & manual CO₂ fire suppression system
Smoke detector with optical & acoustic alarm
Mechanical double door locking (secure containment)
Fire protection in the interior
Exhaust port with reversible overpressure flap
Connections for external inerting (e.g., nitrogen)
Emergency stop / manual activation of the suppression system
Pass-through openings for charging and measuring cables
Result:
The device is designed for tests on lithium-ion cells up to EUCAR Hazard Level 5,
where fire or flames may occur without explosion
or uncontrolled spread.
Not an abuse or destruction testing device, but a safe,
standard-compliant test cabinet for real operational and failure scenarios.
Technical Core Data:
Temperature range: –5 °C to +100 °C
Interior volume: approx. 698 liters
Racks: 4 × stainless steel
Max. load per rack: 45 kg
Max. total load: 100 kg
Electrical Data:
Power supply: 200–240 V AC
Frequency: 50 / 60 Hz
Phase: single-phase
Power consumption: approx. 2.1 kW
Protection: 16 A
Refrigerant: R134a
Protection class: IP 20
Dimensions & Weight:
External (W × H × D): approx. 1250 × 1950 × 900 mm
Internal (W × H × D): approx. 970 × 1250 × 576 mm
In short – why this device is special
true battery safety testing cabinet
compatible with EUCAR Hazard Level 5
integrated CO₂ suppression system
large chamber (≈700 l)
high market & resale attractiveness
Weight: approx. 285–300 kg
Documents & Downloads
Condition: used
Scope of delivery: (See image)
(Changes and errors in technical data are reserved!)
We would be happy to answer any further questions by phone.

Technical characteristic Value
Item ID 24948
Condition Used
Model KB 720 Kühlinkubator
Manufacturer Binder
Weight 300000 g
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[24948].variation.documents[0].data.item.id;

EU Responsible Person

BINDER GmbH
Im Mittleren Ösch 5
78532 Tuttlingen Germany
CustomerSupport@binder-world.com
+49 7462/2005-0

Manufacturer details

Binder
BINDER GmbH
Germany

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system

Climate test cabinets