Vötsch Climate Test Chamber VCL 4330 MH/S with Sunlight Simulation 3300 Liters with Humidity


Item ID: 24415

Condition: Used

Stock: Only ${ $store.getters['24415/currentItemVariation'].stock.net } in stock

The item is already sold

* Excl. VAT excl. Shipping

Vötsch


Type: VCL 4330 MH/S
Climate Test Chamber VCL 4330 MH/S with Sunlight Simulation Water-Cooled
4 x SOL 2000 IC for Sunlight Simulation
Room Volume 3300 Liters
Specifications for Temperature Tests without Irradiation:
Temperature Range without Insulation Lid -30°C to +100°C
with Insulation Lid -40°C to +180°C
Temperature Deviation over Time
in Steady State ±1K
Heating Rate according to DIN 50011 Part 12 3 K/min
Cooling Rate according to DIN 50011 Part 12 2 K/min down to -30°C
Specifications for Climate Tests without Irradiation:
Temperature Range +10°C to +90°C
Temperature Deviation over Time in Steady State ±0.5 K
Humidity Range 10% to 95% r.H.
Dew Point Range +5°C to +88°C
Humidity Deviation over Time in Steady State ± 1 to 3 % r.H.
Specifications for Temperature Tests with Irradiation:
Temperature Range -10°C to +100°C
Temperature Deviation over Time in Steady State ±1K
Heating Rate according to DIN 50011 Part 12 3 K/min
Cooling Rate 1.5 K/min (down to -10°C)
Irradiation Device with Metal Halide Lamp
Irradiation Device:
Weight approx. 1800 kg {Total Weight}
2 kW Halide Lamps 4 pieces
Irradiance 1000 W/m² ± 5%
Wavelength 280 to 3000 nm
Distance from Bottom Edge of Irradiation Unit to Reference Plane 750mm
Size of Reference Plane 1700 x 800 mm
Nominal Voltage 400 V, +6/-10%, 3/N, 50 Hz
Nominal Power 34 kW {Total Connection}
Nominal Current 79 A {Total Connection}
Fuse on Site 100 A (Total Connection)
Specifications for Climate Tests with Irradiation:
Temperature Range +10°C to +80°C
Temperature Deviation over Time in Steady State ±1K
Humidity Range 10% to 85% r.H.
Dew Point Range +5°C to +76°C
Humidity Deviation over Time in Steady State ±1 to 3% r.H.
Sunlight Simulation System:
4 x SOL 2000 IC for Sunlight Simulation
Function Description:
The sunlight simulation system consists of two main parts.
Radiation Part
Control Cabinet
The radiation part consists of 4 radiation modules SOL 2000.
Control Cabinet
The electrical supply for the 4 SOL 2000 modules is provided through the control cabinet,
which houses the control modules, the ballasts, and the
compensation capacitors.
The control cabinet contains all necessary monitoring
and control functions for the operation of the system.
A 3-phase alternating current supply is required.
The entire system is protected by three NH fuses.
A potential-free contact is available as a fault notification in case of lamp failure
or overheating in a lamp module.
Additionally, a potential-free contact is available as a notification
that a lamp is outside the control range.
All operating and display elements are located on the front of the control cabinet door.
Technical Data:
        
Electrical Data    
Complete System
Nominal Voltage: 230 / 400V 3N PE
Nominal Frequency: 50Hz
Nominal Power: 10 kVA compensated
Protection Class: without moisture protection
Radiation Module SOL 2000
Nominal Voltage: 400V
Nominal Frequency: 50Hz
Nominal Current: 6 A compensated
Inrush Current: 10 A compensated
Protection Class: IP20    
Operating Part:
    Nominal Voltage: 230 / 400V 3NPE
Nominal Current: (Continuous) 17 A/ Phase
(Start) approx. 30 A / Phase
Short-Circuit Strength: 6 kA
Control Voltage: 24 V DC
Protection Class: IP 54
built according to VDE: 0113        
Electrical Data:
Climate Test Chamber
Operating Conditions of the Radiation Modules
Humidity:
max. 90% relative humidity at 20°C
max. 50% relative humidity at 35°C
Ambient Temperature:  max. 35°C
Dimensions of the Test Chamber: Width 2000 mm x Depth 1150 mm x Height 1510 mm
External Dimensions: Width 3500 mm x Depth 1450 mm x Height 2100 mm
Refrigerant: R404A
Climate Test Chamber Weight approx. 1550 kg
Irradiation Device Weight approx. 1800 kg
For your safety as a buyer, the following information!
The following points will be carried out in advance on our offered chambers:
1. Function check and replacement of necessary components
2. If necessary, new filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers will undergo a documented test run.
Condition: used
Scope of Delivery: (See Image)
(Changes and errors in the technical data are reserved!)
Further questions can be answered by phone.

Technical characteristic Value
Item ID 24415
Condition Used
Model Vötsch Klimaprüfschrank
Manufacturer Vötsch
Weight 2000000 g
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[24415].variation.documents[0].data.item.id;

EU Responsible Person

Weiss Technik GmbH
Greizer Straße 41-49
35447 Reiskirchen Germany
info@weiss-technik.com
+49 6408 84-0

Manufacturer details

Vötsch
Weiss Technik GmbH
Germany

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system

Climate test cabinets