Vötsch Climate Test Chamber VC 4034 Temperature Range -40°F to +180°F 335 Liters


Item ID: 23405

Condition: Used

Stock: Only ${ $store.getters['23405/currentItemVariation'].stock.net } in stock

The item is already sold

* Excl. VAT excl. Shipping

Vötsch Climate Test Chamber


Here we offer you a climate test chamber.

Air-cooled
Test chamber dimensions: Width 580 x Depth 765 x Height 750 mm
Housing dimensions: Width 780 x Depth 1758 x Height 1800 mm
Key values for temperature tests:
Temperature range: -40 °C to +180 °C
temporal ±0.1 to ±0.5 K
spatial ±0.5 to ±1.5 K
Rate of temperature change according to IEC 60068-3-5
Heating: 3.2 K/min.
Cooling: 3.5 K/min.
Key values for climate tests:
Temperature range: -40 °C to +180 °C
temporal ±0.1 to ±0.3 K
spatial ±0.5 to ±1.0 K
Humidity range: 10 % to 98 % r. H.
Electrical connection:
Nominal voltage: 3/N/PE AC 400V +6/-10% 50Hz
Nominal power: 4.9 kW
Nominal current: 9 A
Refrigerant: R 404A
Test chamber volume: approx. 335 liters
Weight: approx. 515 kg net
Type: VC 4034
For you as a buyer, the following information for your safety!
 
The following points will be carried out in advance on our offered chambers:
1. Function check and replacement of necessary components
2. New filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers will undergo a test run monitored with calibrated devices.
Condition: used
Scope of delivery: (See image)
(Changes and errors in the technical data are reserved!)
 
We would be happy to answer any further questions by phone.

Technical characteristic Value
Item ID 23405
Condition Used
Model Heraeus Vötsch VC 4034
Manufacturer Vötsch
Weight 515000 g
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[23405].variation.documents[0].data.item.id;

EU Responsible Person

Weiss Technik GmbH
Greizer Straße 41-49
35447 Reiskirchen Germany
info@weiss-technik.com
+49 6408 84-0

Manufacturer details

Vötsch
Weiss Technik GmbH
Germany

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system

Climate test cabinets

Show all