CTS Temperature Test Chamber T-55/1000 Li Temperature -50°C to 100°C 1000 liters for testing with lithium-ion batteries


Item ID: 24460

Condition: Used

Stock: Only ${ $store.getters['24460/currentItemVariation'].stock.net } in stock

EUR 28,028.93 *
Content 1 piece
Unit price EUR 28,028.93 / piece
The item is in stock

PAYMENT METHODS:

²


* Excl. VAT excl. Shipping

CTS


CTS Temperature Test Chamber T-55/1000 Li Water-Cooled
Temperature Range: -50°C to +100°C
Nominal Voltage: 400V 3/N 50Hz
Nominal Power: 9.5 kW
Nominal Current: 18 A
Compressor: TFH2511Z / TFH2511Z Refrigerant R404A/R23
The type designation of the CTS cabinet series is composed as follows:
e.g. Type T -55/1000/Li
1) T means: Temperature
2) -55 means: lower temperature -55 °C
3) 1000 means: 1000 liters test chamber volume
4) Li means: special equipment for
tests with lithium-ion batteries
Control:
32 Bit - Controller
LCD Control Panel
Digital display of set and actual values
Numerical input of set values
Error messages in plain text
Ethernet interface
RS 232 interface
Free switching channels for customer-side
test specimen control
Test Chamber Lighting:
energy-efficient, bright LED light strip mounted on the door
Insert Rack:
Stainless steel, support area 970 x 980 mm
Load per rack max. 35 kg
Total load at max. rack loading 100 kg
1a Pass-through NW50
1b Pass-through 2x NW125
1c Pass-through 2x Rp 314°
for Minimax
Technical Performance Data:
Temperature Tests:
Temperature range -50°C to +100°C
Temperature deviation ≤ ±0.3 K over time
Temperature change rate according to IEC 60068-3-5
Heating 2.5 K/min
Cooling 2.5 K/min
Heat Compensation
approx. 4000 W at +20°C
approx. 2200 W at -40°C
The technical data refer to an ambient temperature of +25 °C,
nominal voltage 400V, without test specimen, without heat compensation, and without additional devices.
Options may affect performance data.
Dimensions:
Test volume 1000 liters
Test chamber dimensions height, approx. 1000 mm, width, approx. 860 mm, depth, approx. 1100 mm
External dimensions height, approx. 1950 mm, width, approx. 1270 mm, depth, approx. 1900 mm
Operating Data:
Electrical connection 400 V +6/-10 %, 3/N, 50 Hz
Nominal current 18 A
Nominal power 9.5 kW
Protection class IP 54
Sound pressure level < 58 dB(A)
(according to DIN EN ISO 3744
measured at 1m distance from the front)
Weight approx. 1000 kg
Heat output to the installation room 10.4 kW
Cooling Unit:
water-cooled, low noise, fully hermetic
continuously variable power adjustment through electronic monitoring and control system,
environmentally friendly refrigerants R404A/R23
Additional Equipment:
Description of additional equipment:
1. Test chamber in media-resistant design
The pass-throughs in the test chamber are tightly welded.
2. Fluorosilicone
The door seal consists of a particularly durable fluorosilicone profile.
3. CO Sensor
The CO sensor is used for fire detection in the test chamber. The CO emission in the test chamber is monitored.
A pump is integrated into the gas warning device. The pump draws the gas mixture from the test chamber to the sensor and returns it to the test chamber.
The gas warning device also includes a flow monitoring.
Response time of the sensor min. 60 sec.
The sensor type Dräger CO-LH used is relatively insensitive to H2.
At 1000ppm H₂, only 15ppm CO is displayed.
Threshold 1:
50ppm; as a fault message - light flash active
- door locking active
- temperature control active
- evaluation customer-side
Threshold 2:
200ppm; as a fault message
- light flash and horn active
- device is stopped, temperature control and recirculation fan not active - door locking not active
- evaluation customer-side
H₂ Sensor
The sensor is used to monitor H₂ emissions in the test chamber.
A pump is integrated into the gas warning device.
The pump continuously draws gas from the test chamber and first leads the gas to the sensor of the gas warning system.
Then the gas is returned to the test chamber.
The gas warning device also includes a flow monitoring.
Response time of the sensor min. 60 sec.
Threshold 1:
5% LEL; (≙ 2000ppm) as a fault message - light flash active
- door locking active
- temperature control active
- evaluation customer-side
Threshold 2:
7% LEL; (≙ 3000ppm) as a fault message
- light flash and horn active
- device is stopped, temperature control and recirculation fan not active - door locking not active
- evaluation customer-side
For you as a buyer, the following information for safety!
The following points will be carried out in advance on our offered chambers:
1. Function check and replacement of necessary components
2. If necessary, new filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers will undergo a documented test run.
Condition: used
Scope of delivery: (See image)
(Changes and errors in the technical data are reserved!)
Further questions can be answered by phone.

Technical characteristic Value
Item ID 24460
Condition Used
Model CTS Temperaturprüfschrank
Manufacturer CTS
Weight 1000000 g
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[24460].variation.documents[0].data.item.id;

EU Responsible Person

CTS Clima Temperatur Systeme GmbH
Lotzenäcker 21
72379 Hechingen Germany
info@cts-umweltsimulation.de
+49 7471 - 98 50 - 0

Manufacturer details

CTS
CTS Clima Temperatur Systeme GmbH
Germany

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system