Weiss Temperature Shock Test Chamber ShockEvent T/60/V2/M 60 Liters


Item ID: 22425

Condition: Used

Stock: Only ${ $store.getters['22425/currentItemVariation'].stock.net } in stock

PAYMENT METHODS:

²


* Excl. VAT excl. Shipping

Weiss Temperature Shock Test Chamber


Here we offer you a Weiss Temperature Shock Test Chamber.

Temperature Shock Test Chamber ShockEvent Air Cooling
Temperature Shock Test Chamber ShockEvent T/60/V2 60 Liters
Test chamber dimensions:
External dimensions: Width 900 mm Height 2330 mm Depth 2150 mm
Test chamber dimensions: Width 380 mm Height 370 mm Depth 430 mm
Performance data:
Temperature range warm chamber +50°C to +220°C
Temperature range cold chamber -80°C to +70°C
Heating rate warm chamber 17.0 K/min
Cooling rate cold chamber 3.7 K/min
Heating rate cold chamber 3.2 K/min
Temperature deviation over time ± 0.3 to ± 1.0 K
Temperature homogeneity spatially ± 0.5 to ± 2.0 K
Max. heat compensation cold chamber 2 kW
Calibration value warm chamber 125°C
Calibration value cold chamber -40°C
Switching time between cold and warm chamber <10s
Adjustment time temperature change test <15 min
Consumption and connection data:
Nominal power approx. 8.5 kW
Nominal current approx. 25 A
Protection on site 32 A
Refrigerant R-449a / R-23
Pass-through: Pass-through Ø 80 mm in the ceiling
Control pad:
integrated in the door
Digital setpoint and actual value display of temperatures in warm and cold chamber
Display of the change in operating status
Interfaces:
Ethernet interface 100/1000 Megabit
USB interface
Installation: Size 60 l mobile on wheels
Condenser size 60 l air cooling
Electrical connection:
3/N/PE AC 400 V ±10% 50 Hz
CEE plug with 3.5 m cable
Weight: 800 kg
Type: T/60/V2/M
For you as a buyer, the following information for your safety!
 
The following points will be carried out on our offered chambers in advance:
1. Function check and replacement of necessary components
2. New filling with legally compliant refrigerant
3. After successful inspection, the chambers will undergo a test run monitored with calibrated devices.
Condition: used
Scope of delivery: (See image)
(Changes and errors in the technical data, specifications are subject to change!)
We would be happy to answer any further questions by phone.

document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[22425].variation.documents[0].data.item.id;

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system

Climate test cabinets

Show all