Weiss Temperature Chamber WT1-270/10 Temperature -70°C to 180°C


Item ID: 19196

Condition: Used

Stock: Only ${ $store.getters['19196/currentItemVariation'].stock.net } in stock

The item is already sold

* Excl. VAT excl. Shipping

Weiss Temperature Chamber


Here we offer a Weiss temperature chamber.
Weiss Temperature Chamber water-cooled

with a temperature change rate of 10 K/min

Temperature range: from -70°C to 180°C

Cooling temperature change rate 14.5 K/min

Heating temperature change rate 10.0 K/min

Temperature deviation over time ± 0.1 to ± 0.5 K

Spatial temperature homogeneity ± 0.5 to ± 2.0 K

Test chamber dimensions: Width 580 mm x Depth 615 mm x Height 750 mm

Housing dimensions:
Width 895 mm x Depth 2080 mm x Height 1950 mm
Voltage: 3/N/PE AC 400V±10% 50Hz
Rated power 14 kW Rated current 26 A
Refrigerant: R404A
Weight: 425 kg
Type: WT1-270/70/10
For you as a buyer, the following information for your safety!
 
The following points will be carried out on our offered chambers in advance:
1. Function check and replacement of necessary components
2. New filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers will undergo a test run monitored with calibrated devices.
Condition: used
Scope of delivery: (See image)
(Changes and errors in the technical data, specifications are subject to change!)
 
We would be happy to answer any further questions by phone.

Technical characteristic Value
Item ID 19196
Condition Used
Model WT1-270/70/10
Manufacturer Weiss
Weight 500000 g
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[19196].variation.documents[0].data.item.id;

EU Responsible Person

Weiss Technik GmbH
Greizer Straße 41-49
35447 Reiskirchen Germany
info@weiss-technik.com
+49 6408 84-0

Manufacturer details

Weiss
Weiss Technik GmbH
Germany

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system

Climate test cabinets