Weiss Technik Climate Test Chamber WK 5/55-120 | -55°C to +120°C | 5 m³ | Year of manufacture 2016
Item ID: 25062
Condition: Used
Stock: Only ${ $store.getters['25062/currentItemVariation'].stock.net } in stock
PAYMENT METHODS:
²
* Excl. VAT excl. Shipping
Weiss
Type: WK 5/55-120
Weiss Technik Climate Test Chamber WK 5/55-120
-55°C to +120°C / 5 m³ | Year of manufacture: 2016
The system is water-cooled, powerful, and designed for
professional continuous use.
For sale is a professional climate test chamber from Weiss Technik,
Type WK 5/55-120 from the year of manufacture 2016.
The system is ideally suited for temperature, climate, and material testing in industry,
research, and development.
With a temperature range of -55°C to +120°C and a humidity range of 10–95 % r.H.,
the test chamber covers a wide range of applications.
The spacious test chamber of approximately 5 m³ also allows for testing larger components or assemblies.
Technical Highlights:
Temperature range: -55°C to +120°C
Humidity range: 10–95 % r.H.
Heating/Cooling rate approx. 1.5 K/min
Water cooling required
Technical Data:
Manufacturer: Weiss Technik GmbH
Type: WK 5/55-120
Year of manufacture: 2016
Dimensions / Volume
Test chamber volume: approx. 5 m³
External dimensions (W×D×H): approx. 1900 × 4410 × 2621 mm
Internal dimensions (W×D×H): approx. 1600 × 1700 × 2000 mm
Temperature range
-55 °C to +120 °C
Heating/Cooling rate: approx. 1.5 K/min
Humidity range 10 – 95 % r.H.
Dew point range: +4 °C to +85 °C
Electrical Data:
Power connection: 3/N/PE AC 400V ±10% 50Hz
Rated power: approx. 47 kW (depending on equipment)
Rated current: 71 A
Customer-side protection: 63 A
Cooling water
Water cooling required
Cooling water consumption max. approx. 7 m³/h
Inlet temperature approx. 20 °C
Noise level approx. 76 dB(A)
Weight approx. 4,600 kg
Scope of delivery:
Climate test chamber Weiss Technik WK 5/55-120
Control system
Operating manual (German)
Type plate & documentation
Condition: used
Scope of delivery: (See image)
(Changes and errors in the technical data are reserved!)
Further questions can be answered by phone.
Switch to environmentally friendly refrigerant
The following work can be performed on climate chambers. Please contact us if you are interested.
- Extraction and proper disposal of the previous refrigerant
- Replacement of necessary components, if unavoidable
- Evacuate the circuit and refill it with R 452A, for example.
- Checking / adjusting overheating in expansion valves
- Leak test
- Test run of the system