Weiss Climate Test Chamber WK3-600/0-M 600 liters with humidity -5°C to +90°C
Item ID: 24892
Condition: Used
Stock: Only ${ $store.getters['24892/currentItemVariation'].stock.net } in stock
PAYMENT METHODS:
²
* Excl. VAT excl. Shipping
Weiss
Type: WK3 600/0-M
Weiss Climate Test Chamber Water-Cooled
Type Series WK3-../0
with SIMPAC* Control
Test Chamber Volume: approx. 600 liters
Test Chamber Dimensions: Height 950 mm Width 800 mm Depth 800 mm
External Dimensions: Height 1995 mm Width 1090 mm Depth 1660 mm
Operating Data:
Test Chamber Lighting: LED 24 V, 6.8 W / Halogen Bulb 2 x 24V, 50W
Nominal Voltage: 1/N / PE AC 230 V ±10% 50 Hz
Nominal Power: 2.3 kW
Nominal Current: 10 A
Fuse on site: 16 A slow
Protection Class of Control Cabinet and Operating Panel IP 54
Max. Heat Output to Environment 1200 W
Sound Pressure Level approx.
measured at 1m distance from the front. 1.6m height 46 dB(A)
Data for Temperature Tests:
Temperature Range: -5°C to +90°C
Temperature Deviation over Time in Center of Usable Space ± 0.1 to ± 0.5 K
Spatial Temperature Deviation ± 0.5 to ± 1 K
Temperature Gradient
(according to IEC 60068-3-5)
Rate of Temperature Change
Heating 0.6 K/min
Cooling 0.3 K/min
Heat Compensation at +20 °C to +50 °C 200 W
Data for Climate Tests:
Temperature Range +10°C to +90°C
Dew Point Range 4 to +89.5 °C
Humidity Range 10 to 98 % r.H.
Temporal Temperature Deviation in Center of Usable Space ± 0.1 to ± 0.3 K
Spatial Temperature Deviation ± 0.5 to ± 1 K
Temporal Humidity Deviation + 1 to + 3 % r.H.
Refrigerant: R-404A
Weight: 610 kg
For your safety as a buyer, the following information!
The following points will be carried out on our offered chambers in advance:
1. Function check and replacement of necessary components
2. New filling with legally compliant refrigerant
3. After successful inspection, the chambers will undergo a test run monitored with calibrated devices.
Condition: used
Scope of Delivery: (See Image)
(Changes and errors in the technical data are reserved!)
We are happy to answer any further questions by phone.
Switch to environmentally friendly refrigerant
The following work can be performed on climate chambers. Please contact us if you are interested.
- Extraction and proper disposal of the previous refrigerant
- Replacement of necessary components, if unavoidable
- Evacuate the circuit and refill it with R 452A, for example.
- Checking / adjusting overheating in expansion valves
- Leak test
- Test run of the system