Weiss Climate Cabinet WKL 34/40 -40°C to 150°C with Humidity 34 Liters


Item ID: 24870

Condition: Used

Stock: Only ${ $store.getters['24870/currentItemVariation'].stock.net } in stock

The item is already sold

* Excl. VAT excl. Shipping

White


Type: WKL 34/40
White Climate Chamber Air-Cooled
-40°C to 150°C
Room Volume: 34 Liters
Standard equipment includes:
R1 NW 50 mm
L1 NW 120 mm
Internal Dimensions: Height 310 mm Width 324 mm Depth 300 mm
External Dimensions: Height 1740 mm Width 630 mm Depth 970 mm
Specifications for Temperature Tests:
Temperature Range -40 to +150 °C
Temperature differences in steady state, according to IEC 60068-3-5)
Temperature fluctuation over time in the center of the usable space ± 0.3 to ± 1 K ± 1 to ± 2 K
Spatial temperature deviation (corresponds to temperature gradient) 2 to 4
Rate of temperature change (according to IEC 60068-3-5)
Heating 2.5 K/min
Cooling 3.5 K/min
Heat compensation max. 800 W
Specifications for Climate Tests:
Temperature Range +10 to +95 °C
Humidity Range 10 to 98 % r.H.
Temperature and humidity differences (in steady state, according to IEC 60068-3-5)
Temperature fluctuation over time in the center of the usable space ± 0.3 to ± 0.5 K
Spatial temperature deviation ± 0.5 to ± 1.5
(corresponds to temperature gradient) 1 to 3
Humidity fluctuation over time in the center of the usable space ± 3 to ± 5 % r.H.
Operating Data:
Test chamber lighting Halogen bulb 12V, 20W
Nominal voltage 1/N / PE AC 230 V ± 10 % 50 Hz
Nominal power 1.8 kW
Nominal current 8.0 A
Fuse on-site 16A slow-blow
Heat output air-cooled test chambers
Max. heat output to environment 800 W
Refrigerant: R-404A
Weight: 110 kg
Important Note:
“The table shown in the photos is not included in the sale.”
For your safety as a buyer, the following information!
The following points will be carried out on our offered chambers in advance:
1. Function check and replacement of necessary components
2. If necessary, new filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers will undergo a documented test run.
Condition: used
Scope of delivery: (See image)
(Changes and errors in the technical data are reserved!)
We would be happy to answer any further questions by phone.

Technical characteristic Value
Item ID 24870
Condition Used
Model Weiss Klimaprüfschrank
Manufacturer Weiss
Weight 130000 g
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[24870].variation.documents[0].data.item.id;

EU Responsible Person

Weiss Technik GmbH
Greizer Straße 41-49
35447 Reiskirchen Germany
info@weiss-technik.com
+49 6408 84-0

Manufacturer details

Weiss
Weiss Technik GmbH
Germany

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system

Climate test cabinets

Show all