Weiss ATLAS Solar Climate Test Chamber SC 1000 MHG


Item ID: 24303

Condition: Used

Stock: Only ${ $store.getters['24303/currentItemVariation'].stock.net } in stock

The item is already sold

* Excl. VAT excl. Shipping

Atlas


Type: SC 1000 MHG
Atlas-Solar Climate Test Chamber
ATLAS-Solar Climate Test Chamber Water-Cooled
Refrigerant converted to R452
ATLAS - SolarClimatic
Type: SC 1000 MHG
Technical Features:
Specifications
Test chamber volume 1000 l
Test chamber dimensions Width 1100 mm Depth 950 mm Height 950 mm
External dimensions Width 1390 mm Depth 1850 mm Height 2550 mm
Irradiation:
Radiation source MHG (Metal Halide Global)
1 x 2500 W
Irradiance 800 - 1200 (W/m²)
Spectrum 280 - 3000 nm
Irradiated area 5600 (cm²)
Performance range for temperature tests
Temperature range with irradiation -20 to +100°C
Temperature deviation with irradiation over time ± 1.0 K
Temperature range without irradiation -30 to +100°C
Temperature deviation without irradiation over time 0.1 to 0.5 K
Change rate without irradiation
- Heating according to IEC 4.0 K/min
- Cooling according to IEC 2.5 K/min
Calibration values +23°C and + 80°C without irradiation
Performance range for climate tests:
Temperature range with irradiation +15 to +80°C
Temperature deviation with irradiation over time + 0.5 K
Humidity range with irradiation 10 to 80 % r.H.
Humidity deviation with irradiation over time ±3 to ±5 % r.H.
Dew point range with irradiation +5 to +74°C
Temperature range without irradiation over time ±0.1 to ±0.3
Humidity range without irradiation 10 to 90 % r.H.
Humidity deviation without irradiation over time ±1 to ±3
Dew point range without irradiation +5 to +87°C
Calibration values (without irradiation) +23°C / 50% r.H. and +80°C / 50% r.H
Technical Data:
Electrical connection 3/N/PE AC, 400 V±10%, 50 Hz
Max. connection power 18 kW
Max. current consumption 32.5 A
Humidification water demineralized on-site
Weight 775kg
For your safety as a buyer, the following information!
The following points will be carried out on our offered chambers in advance:
1. Function check and replacement of necessary components
2. If necessary, new filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers will undergo a documented test run.
Condition: used
Scope of delivery: (See image)
(Changes and errors in the technical data are reserved!)
Further questions can be answered by phone.

Technical characteristic Value
Item ID 24303
Condition Used
Model Weiss ATLAS SC 1000 MHG
Manufacturer Weiss
Weight 1700000 g
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[24303].variation.documents[0].data.item.id;

EU Responsible Person

Weiss Technik GmbH
Greizer Straße 41-49
35447 Reiskirchen Germany
info@weiss-technik.com
+49 6408 84-0

Manufacturer details

Weiss
Weiss Technik GmbH
Germany

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system

Climate test cabinets