Vötsch Temperature Testing Chamber with Spraying Device -30 to 80°C 500 Liters VBT 03/500


Item ID: 24597

Condition: Used

Stock: Only ${ $store.getters['24597/currentItemVariation'].stock.net } in stock

EUR 20,652.89 *
Content 1 piece
Unit price EUR 20,652.89 / piece
The item is in stock

PAYMENT METHODS:

²


* Excl. VAT excl. Shipping

Vötsch


Type: VBT 03/500
Air-cooled
Temperature testing chamber with spraying system
Technical Description:
Data for temperature tests without water:
Temperature range -30°C to +80°C
Temperature deviation in steady state
according to IEC 60068-3-5 depending on the test room temperature
temporal ± 1 K
Rate of temperature change according to IEC 60068-3-5
Heating: approx. 2 K/min.
Cooling: approx. 2 K/min.
Data for temperature tests with water
Temperature range +5°C to +80°C
Possible water level -10 mm to 540 mm
Water level = -10 mm
The water level is below the insert grid
Water level = 0 mm
The water level is at the upper edge of the insert grid
Data for standard spraying system
Number of spray nozzles 6
Spray amount per nozzle 0.5 l/min
Data for spraying system according to EN 539-2, Method B
Number of spray nozzles 6
Spray amount per nozzle 4.5 l/min
Additional flooding nozzles 2
Spray amount additional nozzle 6 l/min
Compressed air consumption approx. 3 m³/h
Data for water tank with preheating
Volume approx. 2000 l
Temperature range for preheating +5°C to +40°C
Duration of temperature change
Cooling from +18°C to +5°C approx. 15 h
Heating from +15°C to +40°C approx. 10 h
Data for installation and operation:
Test room: Width 1120 mm Depth 840 mm Height 540 mm
Housing: Width 2700 mm Depth 1300 mm Height 1575 mm
Water tank: Width 1100 mm Depth 800 mm Height 2200 mm
Pass-throughs
1 pass-through ∅ 50 mm installed on the left
Test room volume: approx. 500 liters
Refrigerant R-404a
Electrical connection:
Nominal voltage 3/N/PE AC 400V ± 10% 50Hz
Nominal power approx. 26 kW
Nominal current approx. 44.0 A (neutral conductor loaded)
Protection 63 A gG, site-specific
Weight approx. 890 kg net
For you as a buyer, the following information for your safety!
The following points will be carried out in advance on our offered chambers:
1. Function check and replacement of necessary components
2. If necessary, new filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers will undergo a documented test run.
Condition: used
Scope of delivery: (See image)
(Changes and errors in the technical data, specifications are reserved!)
We would be happy to answer any further questions by phone.

Technical characteristic Value
Item ID 24597
Condition Used
Model Vötsch VBT 03/500
Manufacturer Vötsch
Weight 1200000 g
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[24597].variation.documents[0].data.item.id;

EU Responsible Person

Weiss Technik GmbH
Greizer Straße 41-49
35447 Reiskirchen Germany
info@weiss-technik.com
+49 6408 84-0

Manufacturer details

Vötsch
Weiss Technik GmbH
Germany

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system

Climate test cabinets