Vötsch Temperature and Climate Test Chamber VCL 0010 +10°C…up to +150°C (normal +180°C)


Item ID: 18131

Condition: Used

Stock: Only ${ $store.getters['18131/currentItemVariation'].stock.net } in stock

The item is already sold

* Excl. VAT excl. Shipping

Vötsch Temperature and Climate Test Chamber


Here we offer you a Vötsch Temperature and Climate Test Chamber.
Important:
The chamber currently has a programmed temperature range of +10°C to +150°C (nominal +180°C)
The temperature changes can be made.
If this is desired, please discuss it with us in advance, as changes of this kind involve costs.”
Technical Data:
General Data:
Temperature Test Chamber Climate Test Chamber VCL 0010
Test room volume approx. 100 liters
Weight 170 kg
Data for Mechanical Load
Maximum load (evenly distributed over the surface)
of the test room floor 10 kg
per shelf 10 kg
Total load of the shelf supports 50 kg
Data for Operation
Test room lighting Halogen bulb 12V, 20W
Nominal voltage 1/N / PE AC 230 V ± 10 % 50 Hz
Nominal power 2.7 kW
Nominal current 11.7 A
Fuse on site 16A slow
Protection class control cabinet IP 22
Protection class control panel IP 54
Heat output air-cooled test chambers
Max. heat output to environment 1700 W
Noise Measurement
Sound pressure level measured at 2m distance from the front,
1m height and free field measurement approx. 56 dB(A)
Key Values for Temperature Tests
Temperature range +10 to +180 °C
Temperature fluctuation over time in the center of the usable space ± 0.3 to ± 1 K
Temperature deviation spatially ± 1 to ± 2 K
Temperature gradient 2 to 4 K
Temperature change rate (according to IEC 60068-3-5)
Heating 3 K/min
Heat compensation -
Key Values for Climate Tests
Data for Humidification System
Contents of the storage container approx. 13 l
Water quality demineralized
pH value 6–7
Conductivity 5 - 20 μS/cm
Water consumption
at constant temperature +40 °C, 92 % r.H. approx. 2 l /24 h
Temperature range +10 to +95 °C
Humidity range 10 to 98 % r.H.
Temperature fluctuation over time in
the center of the usable space ± 0.3 to ± 0.5 K
Temperature deviation spatially ± 0.5 to ± 1.5
Temperature gradient 1 to 3 K
(according to IEC 60068-3-5)
Humidity fluctuation over time in the center of the usable space ± 3 to ± 5 % r.H.

Internal dimensions:     Width 490 mm, Height 500 mm, Depth 400 mm
External dimensions: Width 780 mm, Height 1190 mm, Depth 1105 mm
Weight: 510 kg
Refrigerant: R 404A
Type: VCL 0010
For you as a buyer, the following information for your safety!
 
The following points will be carried out on our offered chambers in advance:
1. Function check and replacement of necessary components
2. New filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers will undergo a test run monitored with calibrated devices.
Condition: used
Scope of delivery: (See image)
(Changes and errors in the technical data are reserved!)
 
We are happy to answer any further questions by phone.

Technical characteristic Value
Item ID 18131
Condition Used
Model Vötsch Temperatur und Klimaprüfschrank VCL 0010
Manufacturer Vötsch
Weight 200000 g
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[18131].variation.documents[0].data.item.id;

EU Responsible Person

Weiss Technik GmbH
Greizer Straße 41-49
35447 Reiskirchen Germany
info@weiss-technik.com
+49 6408 84-0

Manufacturer details

Vötsch
Weiss Technik GmbH
Germany

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system

Climate test cabinets