Vötsch Climate Test Chamber with Humidity -70 to 180°C 335 Liters VC 7034


Item ID: 24867

Condition: Used

Stock: Only ${ $store.getters['24867/currentItemVariation'].stock.net } in stock

EUR 15,500.00 *
Content 1 piece
Unit price EUR 15,500.00 / piece
The item is in stock

PAYMENT METHODS:

²


* Excl. VAT excl. Shipping

Vötsch


Type: VC 7034
Climate Test Chamber with Humidity and Air-Cooled
Touch Panel
Powerful 32 Bit Control, Monitoring, and Regulation System
Technical Description:
Specifications for Temperature Tests:
Temperature Range -70°C to +180°C
Temperature Deviation in Steady State
according to IEC 60068-3-5 depending on the test room temperature
temporal ±0.1 to ±0.5 K
spatial ±0.5 to ±2.0 K
Rate of Temperature Change according to IEC 60068-3-5
Heating: approx. 3 K/min.
Cooling: approx. 2 K/min.
Heat Compensation max. approx. 1000 W
Specifications for Climate Tests:
Temperature Range +10°C to +95°C
Temperature Deviation in Steady State
according to IEC 60068-3-5 depending on the test room temperature
temporal ±0.1 to ±0.3 K
spatial ±0.5 to ±1.0 K
Distance of Measurement Points from Wall 100 mm
Humidity Range 10% to 98% r. H.
Dew Point Range -3°C to +94°C
Humidity Deviation in Steady State,
depending on the climate value
temporal ±1.0% r. H. to ±3.0% r. H.
Heat Compensation 400 W at 25°C to 90°C
Humidity up to max. 90% r. H. is maintained
Data for Setup and Operation:
Test Room: Width 580 mm Depth 765 mm Height 750 mm
Housing (Dimensions for Transport): Width 865 mm Depth 1595 mm Height 1780 mm
Pass-Throughs
1 Pass-Through ∅ 50 mm installed on the left
1 Pass-Through ∅ 125 mm installed on the left
1 Pass-Through ∅ 125 mm installed on the right
Test Room Volume: approx. 335 liters
Refrigerant R-404a + R23
Electrical Connection:
Nominal Voltage 3/N/PE AC 400V ± 10% 50Hz
Nominal Power approx. 5.5 kW
Nominal Current approx. 16.0 A
Plug Cekon 16 A
Connection Cable approx. 3.5 m
Protection 16 A slow, site-specific
Weight approx. 500 kg net
For your safety as a buyer, the following information!
The following points will be carried out on our offered chambers in advance:
1. Function check and replacement of necessary components
2. If necessary, new filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers will undergo a documented test run.
Condition: used
Scope of Delivery: (See Image)
(Changes and errors in the technical data are reserved!)
We are happy to answer any further questions by phone.

Technical characteristic Value
Item ID 24867
Condition Used
Model Vötsch VC 7034
Manufacturer Vötsch
Weight 500000 g
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[24867].variation.documents[0].data.item.id;

EU Responsible Person

Weiss Technik GmbH
Greizer Straße 41-49
35447 Reiskirchen Germany
info@weiss-technik.com
+49 6408 84-0

Manufacturer details

Vötsch
Weiss Technik GmbH
Germany

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system

Climate test cabinets

Show all