Vötsch Climate Test Chamber with Humidity -70°C to + 180°C 600 Liters VCS 7060-5
Item ID: 22279
Condition: Used
Stock: Only ${ $store.getters['22279/currentItemVariation'].stock.net } in stock
PAYMENT METHODS:
²
* Excl. VAT excl. Shipping
Vötsch Climate Chamber
Here we offer a Vötsch climate testing chamber.
Vötsch Climate Chamber with Humidity Water-Cooled
600 Liters - Test Room
External Dimensions: Width 1000 mm x Depth 1660 mm x Height 1995 mm
InternalDimensions: Width 800 mm x Depth 800 mm x Height 950 mm
Specifications for Temperature Tests
Temperature Range -70°C to +180°C
Temperature Deviation Over Time: ± 0.1 to ± 0.5 K
Spatial Temperature Deviation: ± 0.5 to ± 2 K
Heating Rate 6 K/min
Cooling Rate 6 K/min
Heat Compensation
at 20°C: 5000W
at -20°C: 5000 W
Specifications for Climate Tests
Temperature Range: +10°C to +95°C
Temperature Deviation Over Time: ± 0.1 to ± 0.3 K
Spatial Temperature Deviation: ± 0.5 to ± 1 K
Humidity Range: 10% to 98% r.H.
Humidity Deviation Over Time: ± 1 to 3% r.H.
Heat Compensation
in the Range of +25°C to +90°C
at a relative Humidity of > 90%: 500W
Weight: approx. 675 kg
Rated Voltage: 3/N/PE AC 400V ±10% 50Hz
Rated Power: 16 kW Rated Current: 29 A
Fuse on site: 32 A slow
RS232
Refrigerant R-404A / R-23
Type: VCS 7060-5
For your safety as a buyer, the following information!
The following points will be carried out on our offered chambers in advance:
1. Function check and replacement of necessary components
2. New filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers will undergo a test run monitored with calibrated devices.
Condition: used
Scope of Delivery: (See Image)
(Changes and errors in the technical data are reserved!)
We would be happy to answer any further questions by phone.
Switch to environmentally friendly refrigerant
The following work can be performed on climate chambers. Please contact us if you are interested.
- Extraction and proper disposal of the previous refrigerant
- Replacement of necessary components, if unavoidable
- Evacuate the circuit and refill it with R 452A, for example.
- Checking / adjusting overheating in expansion valves
- Leak test
- Test run of the system