Vötsch Climate Test Chamber with Humidity -70°C to +150°C 600 Liters VCS 7360-S/3
Item ID: 22734
Condition: Used
Stock: Only ${ $store.getters['22734/currentItemVariation'].stock.net } in stock
PAYMENT METHODS:
²
* Excl. VAT excl. Shipping
Vötsch Climate Chamber
Here we offer a Vötsch climate testing chamber.
Vötsch Climate Chamber with Humidity Water-Cooled
3600 Liters - Test Chamber
External Dimensions: Width 3750mm x Depth 1300mm x Height 2130mm
Internal dimensions (Usable): Width 1970mm x Depth 1100mm x Height 1500mm
Key Values for Temperature Tests
Temperature Range -70°C to +150°C
Temperature Deviation Over Time: ± 0.1 to ± 0.5 K
Spatial Temperature Deviation: ± 0.5 to ± 2 K
Heating Rate 3 K/min
Cooling Rate 3 K/min
Heat Compensation
at 20°C: 8000W
at -20°C: 8000 W
Key Values for Climate Tests
Temperature Range: +10°C to +90°C
Temperature Deviation Over Time: ± 0.1 to ± 0.5 K
Spatial Temperature Deviation: ± 0.1 to ± 1.5 K
Humidity Range: 10% to 95% r.H.
Humidity Deviation Over Time: ± 3 to 5% r.H.
Heat Compensation
in the Range of +25°C to +90°C
at a relative humidity of > 90%: 500W
Weight: approx. 1990 kg
Rated Voltage: 3/N/PE AC 400V ±10% 50Hz
Rated Power: 28kW Rated Current: 46A
Fuse on site: 63 A slow
Refrigerant R-452A / R-23
Type: VCS 7360-S3
from 07.2018
For your safety as a buyer, the following information!
The following points will be carried out on our offered chambers in advance:
1. Function check and replacement of necessary components
2. New filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers will undergo a test run monitored with calibrated devices.
Condition: used
Scope of delivery: (See image)
(Changes and errors in the technical data are reserved!)
Further questions can be answered by phone.
Switch to environmentally friendly refrigerant
The following work can be performed on climate chambers. Please contact us if you are interested.
- Extraction and proper disposal of the previous refrigerant
- Replacement of necessary components, if unavoidable
- Evacuate the circuit and refill it with R 452A, for example.
- Checking / adjusting overheating in expansion valves
- Leak test