Vötsch Climate Test Chamber VCS 7130-15 -70°C to +180°C with Humidity 1300 Liters


Item ID: 24666

Condition: Used

Stock: Only ${ $store.getters['24666/currentItemVariation'].stock.net } in stock

EUR 28,000.00 *
Content 1 piece
Unit price EUR 28,000.00 / piece
The item is in stock

PAYMENT METHODS:

²


* Excl. VAT excl. Shipping

Vötsch


Type: VCS 7130-15
Vötsch Climate Test Chamber Water-Cooled
Data for Temperature Tests:
Temperature Range -70°C to 180°C
Temporal Temperature Deviation in Center of Usable Space ± 0.3 to ± 0.8 K
Spatial Temperature Deviation ± 0.5 to ± 2 K
Temperature Gradient 1 to 4 K
Rate of Temperature Change
Heating 16 K/min
Cooling 14.5 K/min
Data for Climate Tests:
Temperature Range +10°C to +95°C
Humidity Range 10 to 98 %r.H.
Temporal Temperature Deviation in Center of Usable Space ± 0.1 to ± 0.3 K
Spatial Temperature Deviation ± 0.5 to ± 1 K
Temperature Gradient 1 to 2 K
Temporal Humidity Fluctuation in Center of Usable Space ± 1 to ± 3 %r.H.
Heat Compensation max. 8000W
Data for Housing Dimensions:
External Dimensions with Hinges and Door Lock
Height 1995 mm, Width 1680 mm, Depth 1855 mm
External Dimensions without Hinges and Door Lock
Height 1995 mm, Width 1300 mm, Depth 1855 mm
Internal Dimensions Height 950 mm, Width 1100 mm, Depth 950 mm
Test Chamber Volume: approx. 1300 Liters
Pass-Throughs:
The test chamber is equipped with pass-throughs on the right and left sides.
These allow measurement lines and testing devices to be placed into the test chamber.
1 X R1 NW 125 mm
1 X R1 NW 140 mm
1 X L1 NW 50 mm
Test Chamber Lighting
Halogen Bulb 24 V, 50W
3/N/PE AC 400V ±10% 50Hz
Rated Power: 44.0 kW
Rated Current: 63.0 A
Fuse on site 100 A slow-blow
max. Heat Output to Environment: 2.2 kW
Permissible Test Material Weight
Total Load of Shelf Supports 100 kg
Test Chamber Floor 150 kg, Insert Shelf 50 kg
Refrigerant Climate R 404A/R 23
Weight: 1900 kg
For your safety as a buyer, the following information!
The following points will be carried out on our offered chambers in advance:
1. Function check and replacement of necessary components
2. If necessary, new filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers will undergo a documented test run.
Condition: used
Scope of Delivery: (See Image)
(Changes and errors in technical data are reserved!)
We would be happy to answer any further questions by phone.

Technical characteristic Value
Item ID 24666
Condition Used
Model Vötsch VCS 7130-15
Manufacturer Vötsch
Weight 1900000 g
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[24666].variation.documents[0].data.item.id;

EU Responsible Person

Weiss Technik GmbH
Greizer Straße 41-49
35447 Reiskirchen Germany
info@weiss-technik.com
+49 6408 84-0

Manufacturer details

Vötsch
Weiss Technik GmbH
Germany

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system

Climate test cabinets

Show all