Vötsch Climate Test Chamber VCS 7080-15 -70°C to +180°C 800 Liters 15 K/min


Item ID: 24400

Condition: Used

Stock: Only ${ $store.getters['24400/currentItemVariation'].stock.net } in stock

The item is already sold

* Excl. VAT excl. Shipping

Vötsch


Type: VCS 7080-15
Heraeus Vötsch Climate Test Chamber Water-Cooled
VCS 7080-15  with temperature change rate of 15 K/min
Standard configuration built-in
R1: NW 125 mm
L1: NW 50 mm
  External dimensions: Width 1300 x Depth 2675 x Height 2020 (mm)
Test chamber dimensions Width 1100 x Depth 806 x Height 920 (mm)
Test chamber lighting:
Halogen bulb 24 V, 50W
Nominal voltage: 3/N/PE AC, 400 V ±10 %, 50 Hz
Nominal power: 25 kW Nominal current: 44 A
Fuse, site-specific: 63 A slow-blow
Data for temperature tests:
Temperature range  -70°C to + 180°C
Temperature deviation over time in the center of the usable space ± 0.3 K to ± 0.8 K
Temperature deviation spatially ± 0.5K to ± 2K
Temperature gradient (according to IEC 60068-3-5) 1 to 4 K
Temperature change rate (according to IEC 60068-3-5)
Heating 16 K/min
Cooling 15.5 K/min
Heat compensation at -20°C 8000 W
Heat compensation at +20°C 8000 W
Data for climate tests:
Temperature range  +10°C to + 95°C
Temperature deviation over time in the center of the usable space ± 0.2 K to ± 0.5 K
Temperature deviation spatially ± 0.5 K to ± 1 K
Temperature gradient (according to IEC 60068-3-5) 1 to 2 K
Humidity deviation over time in the center of the usable space ± 1 % to ± 3 % r.H.
Humidity range 10 % to 95 % r.H.
Chamber is water-cooled
Refrigerants R-404A and R-23
Approx. 800 liters
Weight approx. 1800 kg
For your safety as a buyer, the following information!
The following points will be carried out on our offered chambers in advance:
1. Function check and replacement of necessary components
2. If necessary, new filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers will undergo a documented test run.
Condition: used
Scope of delivery: (See image)
(Changes and errors in the technical data are reserved!)
Further questions can be answered by phone.

Technical characteristic Value
Item ID 24400
Condition Used
Model Vötsch Klimaprüfschrank VCS
Manufacturer Heraeus Vötsch
Weight 1800000 g
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[24400].variation.documents[0].data.item.id;

EU Responsible Person

Weiss Technik GmbH
Greizer Straße 41-49
35447 Reiskirchen Germany
info@weiss-technik.com
+49 6408 84-0

Manufacturer details

Heraeus Vötsch
Weiss Technik GmbH
Greizer Straße 41-49
35447 Reiskirchen Germany
info@weiss-technik.com
+49 6408 84-0

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system

Climate test cabinets