Vötsch Climate Test Chamber VC 4034 Temperature Range -40°F to +180°F 335 Liters
Item ID: 23195
Condition: Used
Stock: Only ${ $store.getters['23195/currentItemVariation'].stock.net } in stock
PAYMENT METHODS:
²
* Excl. VAT excl. Shipping
Vötsch Climate Test Chamber
Here we offer you a Vötsch climate test chamber.
Converted to 452A!
Air-cooled
Test chamber dimensions: Width 580 x Depth 765 x Height 750 mm
Housing dimensions: Width 780 x Depth 1758 x Height 1800 mm
Specifications for temperature tests:
Temperature range: -40 °C to +180 °C
temporal ±0.1 to ±0.5 K
spatial ±0.5 to ±2.0 K
Rate of temperature change according to IEC 60068-3-5
Heating: 3.2 K/min.
Cooling: 3.5 K/min.
Specifications for climate tests:
Temperature range: -40 °C to +180 °C
temporal ±0.1 to ±0.3 K
spatial ±0.5 to ±1.0 K
Humidity range: 10% to 98% r. H.
Electrical connection:
Nominal voltage: 3/N/PE AC 400V ±10% 50Hz
Nominal power: 4.1 kW
Nominal current: 12.5 A
Refrigerant: R 404A
Test chamber volume: approx. 335 liters
Weight: approx. 460 kg net
Type: VC 4034
For you as a buyer, the following information for your safety!
The following points will be carried out on our offered chambers in advance:
1. Function check and replacement of necessary components
2. New filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers will undergo a test run monitored by calibrated devices.
Condition: used
Scope of delivery: (See image)
(Changes and errors in the technical data are reserved!)
We would be happy to answer any further questions by phone.
Switch to environmentally friendly refrigerant
The following work can be performed on climate chambers. Please contact us if you are interested.
- Extraction and proper disposal of the previous refrigerant
- Replacement of necessary components, if unavoidable
- Evacuate the circuit and refill it with R 452A, for example.
- Checking / adjusting overheating in expansion valves
- Leak test
- Test run of the system