Vötsch Climate Test Chamber VC³ 4034 R 300 m. Humidity 335 liters Temperature -42°C to +180°C


Item ID: 23958

Condition: Used

Stock: Only ${ $store.getters['23958/currentItemVariation'].stock.net } in stock

The item is already sold

* Excl. VAT excl. Shipping

Heraeus Vötsch


Type: VC³ 4034 R 300
Heraeus Vötsch Climate Test Chamber with Humidity and Air Cooling
The device achieves an air exchange of 300 cubic meters (m³) per hour. (R 300)
The chamber has the option for an external test box with an external sensor connection,
thus also usable as an External Temperature Control Unit.
Weight of the test chamber increases by approximately 20 kg
Nominal power increases by: approximately 0.5 kW
Sound pressure level increases by: approximately 5 dB(A)
Dimensions for connection hoses/pipes
Inner diameter of the nozzles: 150 mm
Outer diameter (flanged): 265 mm
Data for temperature tests
Temperature range: -15°C to +150°C
Temporal temperature deviation: ± 1 K
Recirculation volume: 300 m³/h
Data for climate tests
Temperature range: +10°C to +80°C
Temporal temperature deviation: ± 0.5 K
Humidity range: +10 % r.H. to +80 % r.H.
Temporal humidity deviation: ± 2 % r.H. to ± 5 % r.H.
These specifications refer to tests with a connection of 2 m in a closed air circuit
(direct connection of supply and return air of the test chamber), measured at the end of the connection.
(Converted to refrigerant R-452A)
Control Panel "Touchpanel"
On the control panel "Touchpanel", all control and operation commands can be triggered by simply touching the
function symbols.
Adjustable Feet
To compensate for uneven floors and to ensure adequate
ventilation of the machine part, the test chamber is equipped with adjustable feet.
Pass-Throughs
The test chamber is equipped with pass-throughs on the right and left sides.
Through these, measuring lines
and testing devices can be placed into the test room.
Pipe pass-throughs: approx. 1x 50 mm ∅ 2x 125 mm ∅ and 1x 260 mm ∅
Test room dimensions: Width 580 x Depth 765 x Height 750 mm
Housing dimensions: Width 870 x Depth 1860 x Height 1800 mm
Key figures for temperature tests:
Temperature range: -42°C to +180°C
Temporal ± 0.1 to ± 0.5 K
Spatial ± 0.5 to ± 1.5 K
Temperature change rate according to IEC 60068-3-5
Heating: 3.2 K/min
Cooling: 4 K/min
Key figures for climate tests:
Temperature range: -10°C to +95°C
Dew point range: -3 to +94°C
Humidity range: 10% to 98% r.H.
Temporal temperature deviation
± 0.1 to ± 0.3 K
Spatial ± 0.5 to ± 1 K
Temporal humidity deviation ± 1 to 3 % r.H.
Data for operation:
Halogen lamp 2x 24 V, 50 W / LED strip 24 V, 6.8 W
Nominal voltage: 3/N/PE AC 400V ±10% 50Hz
Nominal power: 4.1 kW Nominal current: 14 A
Protection: 16 A slow, site-specific
Refrigerant: R-452A
Test room volume: approx. 335 liters
Weight: approx. 490 kg net
For your safety as a buyer, the following information!
The following points will be carried out on our offered chambers in advance:
1. Function check and replacement of necessary components
2. If necessary, new filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers will undergo a documented test run.
Condition: used
Scope of delivery: (See image)
(Changes and errors in the technical data are reserved!)
We are happy to answer any further questions by phone.

Technical characteristic Value
Item ID 23958
Condition Used
Model Vötsch VC³4034
Manufacturer Vötsch
Weight 515000 g
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[23958].variation.documents[0].data.item.id;

EU Responsible Person

Weiss Technik GmbH
Greizer Straße 41-49
35447 Reiskirchen Germany
info@weiss-technik.com
+49 6408 84-0

Manufacturer details

Vötsch
Weiss Technik GmbH
Germany

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system

Climate test cabinets

Show all