Vötsch Climate Test Chamber VC 4027/S Temperature Range -40°C to +160°C Test Chamber Explosion Protected


Item ID: 24752

Condition: Used

Stock: Only ${ $store.getters['24752/currentItemVariation'].stock.net } in stock

The item is already sold

* Excl. VAT excl. Shipping

Vötsch


Type: VC 4027/S
Climate Test Chamber air-cooled
Test chamber is explosion-proof. EX protection marking: II 2 G bc II B T3
Test chamber dimensions: Width 580 x Depth 615 x Height 750 mm
Housing dimensions: Width 870 x Depth 1920 x Height 1925 mm
Test chamber volume: approx. 270 liters
Key values for temperature tests:
Temperature range: -10 °C to +160 °C
Temperature deviation in steady state, according to IEC 60068-3-5
depending on the test chamber temperature
temporal ±0.1 to ±0.5 K
spatial ±0.5 to ±2.0 K
Rate of temperature change according to IEC 60068-3-5
Heating: 2 K/min.
Cooling: 2 K/min.
Heat compensation max. 2000 W
Key values for climate tests:
Temperature range: +10 °C to +95 °C
Temperature deviation in steady state, according to IEC 60068-3-5,
depending on the test chamber temperature
temporal ±0.1 to ±0.3 K
spatial ±0.5 to ±1.0 K
Humidity range: 10 % to 98 % r. H.
Pipe connection L1: NW 50 mm
Pipe connection R1: NW 125 mm
Electrical connection:
Nominal voltage: 3/N/PE AC 400V ±10% 50Hz
Nominal power: 11 kW
Nominal current: 23 A
Protection: 32 A slow, site-specific
Weight: approx. 490 kg net
Refrigerant R 404A
For your safety as a buyer, the following information!
The following points will be carried out on our offered chambers in advance:
1. Function check and replacement of necessary components
2. If necessary, new filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers will undergo a documented test run.
Condition: used
Scope of delivery: (See image)
(Changes and errors in the technical data, specifications are subject to change!)
Further questions can be answered by phone.

Technical characteristic Value
Item ID 24752
Condition Used
Model Vötsch Klimaprüfschrank VC 4027/S
Manufacturer Vötsch
Weight 500000 g
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[24752].variation.documents[0].data.item.id;

EU Responsible Person

Weiss Technik GmbH
Greizer Straße 41-49
35447 Reiskirchen Germany
info@weiss-technik.com
+49 6408 84-0

Manufacturer details

Vötsch
Weiss Technik GmbH
Germany

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system

Climate test cabinets