Vötsch Climate Test Chamber Temperature -70°C to +180°C VC 7150


Item ID: 22838

Condition: Used

Stock: Only ${ $store.getters['22838/currentItemVariation'].stock.net } in stock

The item is already sold

* Excl. VAT excl. Shipping

Vötsch


Type: VC 7150
Vötsch Climate Test Chamber Water-Cooled
1578 Liters - Test Space

External Dimensions: Width 1200 mm x Depth 2270 mm x Height 1985 mm
Internal Dimensions: Width 1000 mm x Depth 1025 mm x Height 1540 mm
Specifications for Temperature Tests
Temperature Range -70°C to +180°C
Temporal Temperature Deviation: ± 0.5 K
Spatial Temperature Deviation: ± 1.5 K
Heating Rate according to DIN 50011 Part 12: 2 K/min
Cooling Rate according to DIN 50011 Part 12: 1.5 K/min
Heat Compensation
at 0°C: 1700 W
max. 2000 W
Specifications for Climate Tests
Temperature Range: +10°C to +90°C
Temporal Temperature Deviation: ± 0.3 K
Spatial Temperature Deviation: ± 1 K
Humidity Range: 10% to 98% r.H.
Dew Point Range: +5°C to 89.5°C
Temporal Humidity Deviation: ± 1 to 3% r.H.
Heat Compensation
in the range of +25°C to +90°C,
at a relative humidity of > 90%: 500 W
Water Consumption for Humidification at
constant +40°C and 92% r.H.: < 3 L / 24 h
Data for Installation and Operation
Cooling Water Consumption at Full Load
(Cooling Water Temperature +18°C and Δt=10 K): approx. 1.2 m³/h
Weight: Climate Test Chamber: approx. 1000 kg
Nominal Voltage: 400 V, +6/-10%, 3/N, 50 Hz
Nominal Power: 8.6 kW
Nominal Current: 17 A
Fusing on Site: 32 A slow-blow
RS232
Refrigerant R-404A and R23
For your safety as a buyer, the following information!
The following points will be carried out on our offered chambers in advance:
1. Function check and replacement of necessary components
2. If necessary, new filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers will undergo a documented test run.
Condition: used
Scope of Delivery: (See Image)
(Changes and errors in technical data are reserved!)
We would be happy to answer any further questions by phone.

Technical characteristic Value
Item ID 22838
Condition Used
Model Vötsch VC 7150 70°C bis +180°C
Manufacturer Vötsch
Weight 1400000 g
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[22838].variation.documents[0].data.item.id;

EU Responsible Person

Weiss Technik GmbH
Greizer Straße 41-49
35447 Reiskirchen Germany
info@weiss-technik.com
+49 6408 84-0

Manufacturer details

Vötsch
Weiss Technik GmbH
Germany

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system

Climate test cabinets