Technical Data:
Test Chamber Designation: VCS 4100-5
External dimensions with hinges and door lock
Height 1995 mm, Width 1680 mm, Depth 1855 mm
External dimensions without hinges and door lock
Height 1995 mm, Width 1300 mm, Depth 1855 mm
Test chamber volume: approx. 990 liters
Cooling / Heating Operation -40 to 180°C
Climate Operation 10 – 95°C
Humidity 10...98 % r. F.
Temperature change per time
Cooling down approx. 6.7 K/min
Heating up approx. 6.1 K/min
3/N / PE AC 400 V ± 10% 50 Hz or 3/N / PE AC 440 V ± 10% 60 Hz
Rated power: 23
kW
Rated current: 33 A
Fuse on site: 63 A slow
Max. heat output to environment: 1.7 kW
Permissible test load weight
Total load of shelf supports 100kg, test chamber floor 35kg, insert shelf 50kg
Weight: 840 kg
Type: VCS 4100-5
For your safety as a buyer, the following information!
 
The following points will be carried out on our offered chambers in advance:
1. Function check and replacement of necessary components
2. New filling with legally compliant refrigerant
3. After successful inspection, the chambers will undergo a test run monitored with calibrated devices.
Condition: used
Scope of delivery: (See image)
(Changes and errors in the technical data, specifications are subject to change!)
 

Technical characteristic Value
Item ID 20290
Condition Used
Model Vötsch VCS 4100-5
Manufacturer Vötsch
Weight 950000 g
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[20290].variation.documents[0].data.item.id;

EU Responsible Person

Weiss Technik GmbH
Greizer Straße 41-49
35447 Reiskirchen Germany
info@weiss-technik.com
+49 6408 84-0

Manufacturer details

Vötsch
Weiss Technik GmbH
Germany

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system