Type: PINK VT 300-600-300-1
Vacuum Drying Chamber
Space requirements for operation and maintenance
Width min. 3000 mm Height min. 2500 mm Depth min. 2500 mm
Dimensions VT
External dimensions:
Width approx. 1170 mm Depth approx. 1380 mm Height approx. 1615 mm
Total height 1970 mm
Internal dimensions:
Width approx. 360 mm Depth approx. 735 mm Height approx. 300 mm
Volume of vacuum chamber 79 liters
Operating temperature min. 20°C max. 150°C
Electrical connection connection power 3 x 400 V; 50 Hz N / PE 8.5 kVA
Heating power 6 kW
Cooling power 6 kW at 15°C media temperature and 15°C cooling water
Heating: Floor, ceiling, side walls, and door double-walled, heat transfer medium circulated
Heat transfer medium: Heat transfer oil Mobiltherm 32 Filling quantity approx. 20 liters depending on the length of the piping
Compressed air pressure 6 bar
Nitrogen N2 pressure Nitrogen N2 pressure
Permissible operating overpressure: Vacuum chamber -1 / +0.1 bar
Heating system 3 bar
Weight: 300 kg
Condition: used
Scope of delivery: (PINK VT 300-600-300-1, operating manual)
(Changes and errors in the technical data, specifications are reserved!)
We would be happy to answer any further questions by phone.

Technical characteristic Value
Item ID 16189
Condition Used
Model Vakuumtrockenschrank PINK VT 300-600-300-1
Manufacturer Pink
Weight 600000 g
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[16189].variation.documents[0].data.item.id;

EU Responsible Person

PINK GmbH Thermosysteme
Am Kessler 6
97877 Wertheim Germany
info@pink.de
+49 9342 919-0

Manufacturer details

Pink
Germany

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system