Heraeus Vötsch Temperature and Climate Test Chamber VC 350 LAF/S


Item ID: 18195

Condition: Used

Stock: Only ${ $store.getters['18195/currentItemVariation'].stock.net } in stock

The item is already sold

* Excl. VAT excl. Shipping

Heraeus Vötsch


Type: VC 350 LAF/S
Heraeus Vötsch Temperature and Climate Testing Chamber

Water-Cooled
Heraeus Vötsch Temperature and Climate Testing Chamber with Simcon/32-NET Control
TECHNICAL DATA
General Information:
Test chamber volume: approx. 345 l
Weight: 375 kg
Data for Mechanical Load:
maximum total load of the test chamber floor 60 kg
of which weighing stone 45 kg
Scale + test specimen approx. 15 kg
Storage shelves 3 kg
per shelf permissible load approx. 100 g
Data for Operation:
Interface RS 232
The RS 232 interface is used, for example, for external control via computer.
Lighting Operating Manual Safety Workbench
Interference emission; immunity standards met see Declaration of Conformity
Nominal voltage 3/N / PE AC 400 V ± 10% 50 Hz
Nominal power 5 kW
Nominal current 9 A
Fuse customer-side 16A slow-blow
Noise Measurement:
Sound pressure level measured at 1m distance from the front,
1m height and free field measurement approx. 58 dB(A)
Data for Temperature Tests:
Temperature +20 to +30 °C
Temperature fluctuation over time in the center of the usable space ± 1 K
Heat compensation 100 W
Data for Climate Tests:
Data for Humidification System
Content of the water tank approx. 10 l
Water quality demineralized
pH value 6 – 7
Conductivity max. 10 μs/cm
Temperature +20 to +30 °C
Humidity 30 to 70 % r.H.
Dew point +9.5 °C / +21 °C
Temperature fluctuation over time
in the center of the usable space
(in steady state, according to IEC 60068-3-5)
± 3 K
For your safety as a buyer, the following information!
 
The following points will be carried out on our offered chambers in advance:
1. Function check and replacement of necessary components
2. If necessary, new filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers will undergo a documented test run.
Condition: used
Scope of delivery: (See image)
(Changes and errors in the technical data, specifications are reserved!)
We would be happy to answer any further questions by phone.

Technical characteristic Value
Item ID 18195
Condition Used
Model Heraeus Vötsch Temperatur- und Klimaprüfschrank VC 350 LAF/S
Manufacturer Heraeus Vötsch
Weight 375000 g
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[18195].variation.documents[0].data.item.id;

EU Responsible Person

Weiss Technik GmbH
Greizer Straße 41-49
35447 Reiskirchen Germany
info@weiss-technik.com
+49 6408 84-0

Manufacturer details

Heraeus Vötsch
Weiss Technik GmbH
Greizer Straße 41-49
35447 Reiskirchen Germany
info@weiss-technik.com
+49 6408 84-0

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system

Climate test cabinets