Designation VC³ 4100
With temperature change rate of 3 K/min
Test chamber dimensions: H x W x D, approx. 950 x 1100 x 950 mm
TEMPERATURE TESTS:
Minimum temperature -42°C
Maximum temperature +180°C
Temperature change rate Cooling 3.0 K/min
Temperature change rate Heating 4.0 K/min
Temperature deviation over time ± 0.1 to ± 0.5 K
Temperature homogeneity spatially ± 0.5 to ± 1.5 K
Heat compensation at +20 °C 4500W
CLIMATE TESTS:
Minimum temperature +10°C
Maximum temperature +95°C
Dew point temperature range +4°C to +94°C
Humidity range 10 to 98% r. H.
Humidity deviation over time ± 1 to ± 3% r. H.
Temperature deviation over time ± 0.1 to ± 0.3K
Temperature homogeneity spatially ± 0.5 to ± 1.0K
Heat compensation max. 500W
Voltage: 3/N/PE AC 400V±10% 50Hz
Rated power: 11.5 kW
Rated current: 22 A
Refrigerant: R 404A
Weight approx. 900 kg
Compressor has been renewed!
For your safety as a buyer, the following information!
 
The following points will be carried out on our offered chambers in advance:
1. Function check and replacement of necessary components
2. New filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers will undergo a test run monitored with calibrated devices.
Type: VC³ 4100 / Year of manufacture 2014
Condition: used
Scope of delivery: (See image)
(Changes and errors in the technical data, specifications are reserved!)
 
We would be happy to answer any further questions by phone.

Technical characteristic Value
Item ID 19524
Condition Used
Model Heraeus Vötsch Klimaprüfschrank VC³ 4100
Manufacturer Heraeus Vötsch
Weight 900000 g
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[19524].variation.documents[0].data.item.id;

EU Responsible Person

Weiss Technik GmbH
Greizer Straße 41-49
35447 Reiskirchen Germany
info@weiss-technik.com
+49 6408 84-0

Manufacturer details

Heraeus Vötsch
Weiss Technik GmbH
Greizer Straße 41-49
35447 Reiskirchen Germany
info@weiss-technik.com
+49 6408 84-0

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system