Feutron Climate Test Chamber KPK 200 Type: 3423/16 -40 to 100°C 280 Liters
Item ID: 24004
Condition: Used
Stock: Only ${ $store.getters['24004/currentItemVariation'].stock.net } in stock
PAYMENT METHODS:
²
* Excl. VAT excl. Shipping
Feutron
Type: 3423/16
Climate Test Chamber Air-Cooled
Test Chamber Nominal Volume: 280 Liters
Dimensions:
Outer Chamber: 1500 x 2040 x 1075 mm
Fan Motor and Door Removed: 1500 x 1940 x 925 mm
Test Space: 720 x 690 x 560 mm
Door Opening Width: 720 x 690 mm
Viewing Window, Heated: 500 x 600 mm
Temperature Range:
-40 to +100 °C
Heating: 3.0 K /min on average
Cooling Climate Operation: 0.6 K /min
Temperature Operation: 3.0 K /min on average
Climate Range +10 to +95 °C
Humidity Range: 10 to 95 % r.H.
Humidity Difference, over time: ≤ ± 3.0
Pass-Through Opening 1 piece, Left Side Wall 75 mm ∅
Pass-Through Opening 1 piece, Right Side Wall 75 mm ∅
Water Connection for Humidifier: City Water or Distilled Water 20l Storage Tank
Power Connection: 400 V 3/N/PE, AC, 50 Hz
Maximum Power Requirement: 6.3 kW
Weight: 550 Kg
Refrigerant: R134 a / R507
For your safety as a buyer, the following information!
The following points will be carried out on our offered chambers in advance:
1. Function check and replacement of necessary components
2. If necessary, new filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers will undergo a documented test run.
Condition: used
Scope of Delivery: (See Image)
(Changes and errors in the technical data are reserved!)
We would be happy to answer any further questions by phone.
Switch to environmentally friendly refrigerant
The following work can be performed on climate chambers. Please contact us if you are interested.
- Extraction and proper disposal of the previous refrigerant
- Replacement of necessary components, if unavoidable
- Evacuate the circuit and refill it with R 452A, for example.
- Checking / adjusting overheating in expansion valves
- Leak test
- Test run of the system