Type: 3707/06
Feutron Climate Test Chamber with Solar Simulation
Temperature
Temperature range: -40...90°C
Climate range (without heat load) 10...85°C
Temperature difference, time during temperature operation ≤ ± 0.5 K/min
Temperature difference, time during climate operation ± 1.0... ± 1.5 K/min
Feutron Climate Test Chamber
Here we offer a Feutron Climate Test Chamber.
Feutron Climate Test Chamber
Temperature
Temperature range: -40...90°C
Climate range (without heat load) 10...85°C
Temperature difference, time during temperature operation ≤ ± 0.5 K/min
Temperature difference, time during climate operation ± 1.0... ± 1.5 K/min
Change rates
Heating 2.0 on average K/min
Cooling 2.0 on average K/min
The standard IEC 1646 is realizable with testing, solar modules with a maximum weight of 160kg and nonlinear ramps
(Setpoint jump)
Drying
Air drying condensation drying
Humidity
Air humidification steam humidification
Humidity range 10... 95 % r.H.
Humidity difference, time ≤ ± 3.0... ±5.0 %r.H.
Dew point range 8... 80.9 °C
Test chamber
Test chamber volume 7.0m³
Dimensions
Test chamber dimensions Width 1500 mm x Height 2250 mm x Depth 2000 mm
Swing door DIN right heated, i,l.. Width 900 mm x Height 2100 mm
Viewing window, heated with blackout flap and lockable Width 500 mm x Height 600 mm
Pass-through opening 2x on front side right ∅ 75 mm
Irradiation unit test chamber
Cutout 4x in side wall left for SC 2500 each 850 x 700 mm
Area to be irradiated when using SC 2500 1200 x 1200 mm
Radiation intensity at 1.0 m wall distance 1000 ± 10% W/m²
Lighting 1x Halogen lamp 200W
Cooling system Bitzer compressor semi-hermetic 2-stage, air-cooled
Condenser: Split condenser (outdoor installation winter control)
Refrigerant R507 /R23
Temperature control system
Cooling brine medium Tyfocor / water
Mixing ratio 50/50%
Density of cooling brine 1.072 kg/dm³
Filling volume approx. 80dm³
Pyranometer temperature range -40... 150°C
Max. radiation absorption 4000 W/m²
Spectral range (50% points) 305... 2800nm
Sensitivity (nominal) 7 µV/W/m²
Power supply test chamber
Electrical connection 400V, 3/N/PE, AC, 50Hz
Connection type Fixed connection
Max. simultaneous power requirement approx. 32.0 kW
Supply line fuse 3 x 63, Char C (A)
Power supply SolarConstant
Electrical connection 400V, 3/N/PE, AC, 50Hz
Connection type
Max. simultaneous power requirement approx. 30.0 kW
Supply line fuse 3 x 63, Char C (A)
External dimensions of the test chamber complete with: Equipment room/control cabinet
and control cabinet for SolarConstant 2500
External dimensions equipment room Width 3440 mm x Height 2580 mm x Depth 2950 mm
/Control cabinet Width 2200 mm x Height 1830 mm x Depth 1100 mm
External dimensions condenser, split Width 900 mm x Height 950 mm x Depth 2000 mm
Weight test chamber, total approx. 2700 kg
Weight condenser, empty 153 kg
Weight: 2700 Kg
Type: 3707/06
Condition: used
Scope of delivery: (See image)
(Changes and errors in the technical data are reserved!)
We would be happy to answer any further questions by phone.
Change rates
Heating 2.0 on average K/min
Cooling 2.0 on average K/min
The standard IEC 1646 is realizable with testing, solar modules with a maximum weight of 160kg and nonlinear ramps
(Setpoint jump)
Drying
Air drying condensation drying
Humidity
Air humidification steam humidification
Humidity range 10... 95 % r.H.
Humidity difference, time ≤ ± 3.0... ±5.0 %r.H.
Dew point range 8... 80.9 °C
Test chamber
Test chamber volume 7.0m³
Dimensions
Test chamber dimensions Width 1500 mm x Height 2250 mm x Depth 2000 mm
Swing door DIN right heated, i,l.. Width 900 mm x Height 2100 mm
Viewing window, heated with blackout flap and lockable Width 500 mm x Height 600 mm
Pass-through opening 2x on front side right ∅ 75 mm
Irradiation unit test chamber
Cutout 4x in side wall left for SC 2500 each 850 x 700 mm
Area to be irradiated when using SC 2500 1200 x 1200 mm
Radiation intensity at 1.0 m wall distance 1000 ± 10% W/m²
Lighting 1x Halogen lamp 200W
Cooling system Bitzer compressor semi-hermetic 2-stage, air-cooled
Condenser: Split condenser (outdoor installation winter control)
Refrigerant R507 /R23
Temperature control system
Cooling brine medium Tyfocor / water
Mixing ratio 50/50%
Density of cooling brine 1.072 kg/dm³
Filling volume approx. 80dm³
Pyranometer temperature range -40... 150°C
Max. radiation absorption 4000 W/m²
Spectral range (50% points) 305... 2800nm
Sensitivity (nominal) 7 µV/W/m²
Power supply test chamber
Electrical connection 400V, 3/N/PE, AC, 50Hz
Connection type Fixed connection
Max. simultaneous power requirement approx. 32.0 kW
Supply line fuse 3 x 63, Char C (A)
Power supply SolarConstant
Electrical connection 400V, 3/N/PE, AC, 50Hz
Connection type
Max. simultaneous power requirement approx. 30.0 kW
Supply line fuse 3 x 63, Char C (A)
External dimensions of the test chamber complete with: Equipment room/control cabinet
and control cabinet for SolarConstant 2500
External dimensions equipment room Width 3440 mm x Height 2580 mm x Depth 2950 mm
/Control cabinet Width 2200 mm x Height 1830 mm x Depth 1100 mm
External dimensions condenser, split Width 900 mm x Height 950 mm x Depth 2000 mm
Weight test chamber, total approx. 2700 kg
Weight condenser, empty 153 kg
Weight: 2700 Kg
Type: 3707/06
Condition: used
Scope of delivery: (See image)
(Changes and errors in the technical data are reserved!)
We would be happy to answer any further questions by phone.

Technical characteristic Value
Item ID 23408
Condition Used
Model Feutron Klimaprüfzelle
Manufacturer Feutron
Weight 2700000 g
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[23408].variation.documents[0].data.item.id;

EU Responsible Person

Feutron Klimasimulation GmbH
Am Weberbrunnen 1
07957 Langenwetzendorf Germany
feutron@feutron.de
+49 (0) 36625 500-60

Manufacturer details

Feutron
Germany

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system