Espec Temperature Shock Chamber TSA-101S-W +60 to 200°C and -70 to 0°C
Item ID: 24638
Condition: Used
Stock: Only ${ $store.getters['24638/currentItemVariation'].stock.net } in stock
PAYMENT METHODS:
²
* Excl. VAT excl. Shipping
Espec
Type: Espec Model: TSA-101S-W
Espec Temperature Shock Chamber Water-Cooled
Thermal Shock Environmental Chamber (Cold Load: -65° to 0°C).
Conventional thermal shock chambers with three zones can be quite large, as three separate chambers are used.
The TSA series has only one chamber, significantly reducing its size.
Instead, it circulates cold, hot, or ambient air as needed.
The conditioned air can be pre-cooled or pre-heated beyond the setpoint,
allowing for faster recovery of the desired state after transfer.
Warm Chamber: +60 to 200°C Cold Chamber: -70 to 0°C
Operating Temperature: 0-40°C
System: Two zone or Three zone testing via flap switching
Performance:
Test Range:
High Temperature Exposure Range: +60 to +200°C
Low Temperature Exposure Range: –70 to 0°C
Temperature Fluctuation ±0.5°C
Hot Chamber:
Temperature Heating Time: Ambient temp. to +200°C within 15 min.
Cooling Chamber:
Temperature Pulldown Time: Ambient temp. to –75°C within 45 min.
Temperature Recovery: within 10 min.
Internal Dimensions (W x H x D mm): 650 x 460 x 370
External Dimensions (W x H x D mm): 1550 x 1900 x 1370
Voltage: AC 400 V
Phases: 3~ (Three-phase)
Frequency: 50/60 Hz
Maximum Power Consumption: 50 A
Refrigeration Unit (Refrigeration Motor): 5.5 kW
Heating (Heater): 6 kW
Refrigerant: R404/R23
Weight: approx. 1150 kg
For your safety as a buyer, the following information!
The following points will be carried out on our offered chambers in advance:
1. Function check and replacement of necessary components
2. If necessary, new filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers will undergo a documented test run.
Condition: used
Scope of delivery: (See image)
(Changes and errors in technical data are reserved!)
We are happy to answer any further questions by phone.
Switch to environmentally friendly refrigerant
The following work can be performed on climate chambers. Please contact us if you are interested.
- Extraction and proper disposal of the previous refrigerant
- Replacement of necessary components, if unavoidable
- Evacuate the circuit and refill it with R 452A, for example.
- Checking / adjusting overheating in expansion valves
- Leak test
- Test run of the system