Binder Incubators Classic.Line BF 720 (08-35712) plus 5°C to 100°C


Item ID: 24180

Condition: Used

Stock: Only ${ $store.getters['24180/currentItemVariation'].stock.net } in stock

The item is already sold

* Excl. VAT excl. Shipping

Binder


Type: BF 720
Incubator Classic.Line
Key Features
Temperature range: Room temperature plus 5°C to 100°C
APT.line™ preheating chamber technology
Adjustable fan speed
Adjustable exhaust flap
Controller with time functions
Inner door made of safety glass (ESG)
Temperature selection monitor Class 3.1 (DIN 12880)
with optical and acoustic temperature alarm
Computer interface: RS 422
Designation BF720-230V
Version Standard
Article number 9110-0243
Performance data Temperature
Temperature range 5°C above room temperature to [°C] 100
Spatial temperature deviation at 37°C [± K] 0.4
Temporal temperature deviation at 37°C [± K] 0.1
Heating time to 37°C [min] 17
Recovery time after 30 s door opening at 37°C [min] 4
Electrical data
Nominal voltage [V] 230
Network frequency [Hz] 50/60
Nominal power [kW] 1.25
Device fuse [A] 12.5
Phase (nominal voltage) 1~
Dimensions
Interior volume [L] 720
Net device weight (empty) [kg] 180
Maximum load per shelf [kg] 45
Dimensions - External dimensions
Width net [mm] 1235
Height net [mm] 1530
Depth net [mm] 865
Wall distance back [mm] 160
Wall distance side [mm] 100
Dimensions - Internal dimensions
Width [mm] 1000
Height [mm] 1200
Depth [mm] 600
Dimensions - Doors
Outer doors 2
Environmentally relevant data
Energy consumption at 37°C 80 Wh/h
Condition: used
Scope of delivery: (See image)
(Changes and errors in technical data are reserved!)
We would be happy to answer any further questions by phone.

Technical characteristic Value
Item ID 24180
Condition Used
Model Binder Inkubatoren
Manufacturer Binder
Weight 180000 g
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[24180].variation.documents[0].data.item.id;

EU Responsible Person

BINDER GmbH
Im Mittleren Ösch 5
78532 Tuttlingen Germany
CustomerSupport@binder-world.com
+49 7462/2005-0

Manufacturer details

Binder
BINDER GmbH
Germany

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system

Climate test cabinets