Binder FED 400 Heating Chamber with Forced Circulation


Item ID: 6932

Condition: Used

Stock: Only ${ $store.getters['6932/currentItemVariation'].stock.net } in stock

The item is already sold

* Excl. VAT excl. Shipping

Binder FED 400 Heating Chamber with Forced Circulation


Here we offer a Binder FED 400 heating chamber with forced circulation.
Electronically controlled APT.line™ preheating chamber technology ensures high temperature accuracy and reproducible results
    Temperature range 5 °C above room temperature up to 300 °C
    MS controller with various time functions
    Controller clock functions: delayed ON, delayed OFF, and temperature-dependent delayed OFF
        Digital temperature setting accurate to the degree
    Adjustable fan speed
    Adjustable air exchange via front air flap controller and rear exhaust pipe Ø 50 mm
    Temperature limit switch class 2 (DIN 12880) with optical temperature alarm
    RS 422 interface for communication software APT-COM DataControlSystem, or switchable to output to printer with interface converter RS 232 / RS 422
    Devices up to 115 liters are stackable
    2 shelves, chrome-plated
FED 400     
    External dimensions     
    Width (mm)     1235
    Height (including feet) (mm)     1025
    Depth (mm)     765
    Plus door handle and exhaust pipe (mm)     105
    Wall distance at the back (mm)     100
    Wall distance on the side (mm)     160
    Exhaust pipe (outer Ø mm)     52
    Steam chamber volume (l)     498
    Number of doors (pieces)     2
            
    Internal dimensions     
    Width (mm)     1000
    Height (mm)     800
    Depth (mm)     500
    Internal volume (l)     400
    Shelf grid (No. series/max.)     2 / 10
    Load per grid (kg)     35
    Permissible total load (kg)     90
    Weight (empty) (kg)     145
            
    Temperature data     
    Temperature range approx. 5 °C above room temperature up to (°C)     300
    Spatial temperature deviation     
    at 70 °C (± K)     1
    at 150 °C (± K)     2.5
    at 300 °C (± K)     4.8
    Temporal temperature deviation at 70 °C (± K)     0.3
    Heating time 1)     
    to 70 °C (Min.)     18
    to 150 °C (Min.)     35
    to 250 °C (Min.)     60
    Recovery time after 30 sec door open 1)     
    at 70 °C (Min.)     2
    at 150 °C (Min.)     17
    at 300 °C (Min.)     21
            
    Air exchange data     
    Air exchange     
    at 70 °C (x/h)     17
    at 150 °C (x/h)     18
    at 300 °C (x/h)     16
            
    Electrical data     
    IP protection class according to EN 60529     IP 20
    Nominal voltage (±10 %) 50 / 60 Hz (V)     400, 3N~
    Nominal power (kW)     3.4
    Energy consumption     
    at 70 °C (W)     520
    at 150 °C (W)     1200
    at 300 °C (W)
Type: Binder FED 400
Scope of delivery. (See image)

The device comes from a stock liquidation.

We deliver with express shipping (DHL, UPS, TNT)
 
Further questions can be answered by phone.

Written orders possible via email or fax
 


Technical characteristic Value
Item ID 6932
Condition Used
Model Binder FED 400 Wärmeschrank mit forcierter Umluft
Manufacturer Binder
Weight 145000 g
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[6932].variation.documents[0].data.item.id;

EU Responsible Person

BINDER GmbH
Im Mittleren Ösch 5
78532 Tuttlingen Germany
CustomerSupport@binder-world.com
+49 7462/2005-0

Manufacturer details

Binder
BINDER GmbH
Germany

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system

Climate test cabinets