ACS Angelantoni Climate Chamber -70°C to +180°C 178 Liters CH160 TC
Item ID: 22923
Condition: Used
Stock: Only ${ $store.getters['22923/currentItemVariation'].stock.net } in stock
PAYMENT METHODS:
²
* Excl. VAT excl. Shipping
ACS Angelantoni Climate Chamber
Here we offer you an ACS Angelantoni Climate Chamber.
Angelantoni Climate Chamber Air-Cooled
Usable volume 178 liters
Internal usable dimensions: Width 560 mm x Depth 564 mm x Height 560 mm
External dimensions: Width 760 mm x Depth 1100 mm x Height 1470 mm
Specifications for Temperature Testing
Temperature range -70°C to +180°C
Temperature accuracy over time ±0.25
R.F. range in 10°C<T<80°C min dew point = 5°C 20% ... 95%
Heating rate➀ (from -70°C to +180°C) 2.5°C/min
Cooling rate➀ (from +180°C to -70°C) 1°C/min
Maximum internal power loss (W) at T=-55°C 100
Nominal voltage: 230V +1+N+G 50Hz 10A
Refrigerant: R404 A / R23
Weight: approx. 300 kg without wheels
Type: CH160 TC
For you as a buyer, the following information for your safety!
The following points will be carried out on our offered chambers in advance:
1. Function check and replacement of necessary components
2. New filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers will undergo a test run monitored with calibrated devices.
Condition: used
Scope of delivery: (See image)
(Changes and errors in the technical data, specifications are reserved!)
Further questions can be answered by phone.
Switch to environmentally friendly refrigerant
The following work can be performed on climate chambers. Please contact us if you are interested.
- Extraction and proper disposal of the previous refrigerant
- Replacement of necessary components, if unavoidable
- Evacuate the circuit and refill it with R 452A, for example.
- Checking / adjusting overheating in expansion valves
- Leak test
- Test run of the system