Weiss Corrosion Alternating Test Device Salt Spray Chamber SaltEvent, Type SC-KWT 1000


Item ID: 24285

Condition: Used

Stock: Only ${ $store.getters['24285/currentItemVariation'].stock.net } in stock

The item is already sold

* Excl. VAT excl. Shipping

Weiss


SC/KWT 1000 SALTEVENT
Salt Spray Test Chamber
System Weiss, Type SaltEvent SC/KWT 1000
including extension for
Climatic Change Test Device according to VDA 621-415
Designation: SaltEvent SC/KWT 1000
Test chamber volume 1028 liters
Dimensions of test chamber
Height with hood 1140 mm
Height without hood 720 mm
Width 1670 mm
Width to spray channel 1560 mm
Depth 645 mm
Dimensions of system
Height 1370 mm
Width 2925 mm
Depth 980 mm
Temperature range
Salt spray fog test 5 K above ambient temperature up to +50 °C
Condensate test 5 K above ambient temperature up to +42 °C
Temperature stability, time ±1.0 K
Water consumption
Salt spray fog test 0.4 l/h
Condensate test 34.0 liters
Consumption of spray solution
Salt spray fog test 0.9 l/h
Compressed air consumption during
Salt spray fog test 1.4 m3/h
Ventilation 7.0 m3/h
Calibration values salt spray fog test +35 °C
Test chamber temperature condensate test +40 °C
Calibration value humidifier temperature, pressure humidifier +49 °C
Voltage: 1/N/PE AC 230V ± 10% 50Hz
Rated power 2 kW
Rated current 8.7 A
Weight: 372 kg
Condition: used
Scope of delivery: (See image)
(Changes and errors in technical data, specifications are reserved!)
We would be happy to answer any further questions by phone.

Technical characteristic Value
Item ID 24285
Condition Used
Model Weiss Salzsprüh-Prüfkammer SC/KWT 1000
Manufacturer Weiss
Weight 200000 g
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[24285].variation.documents[0].data.item.id;

EU Responsible Person

Weiss Technik GmbH
Greizer Straße 41-49
35447 Reiskirchen Germany
info@weiss-technik.com
+49 6408 84-0

Manufacturer details

Weiss
Weiss Technik GmbH
Germany

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system