ascott Atmosphere Salt Spray Chamber Corrosion Testing Chamber 20°C to 70°C + Humidity AT1300IP/3 (50)


Item ID: 24500

Condition: Used

Stock: Only ${ $store.getters['24500/currentItemVariation'].stock.net } in stock

The item is already sold

* Excl. VAT excl. Shipping

ascott


Type: ascott AT1300IP/3
Atmosfär Corrosion Test Chamber with Data Logger Transfer Software.
Main Features of Atmosfär
Oscillating spray bar with magnetically coupled drive system.
Positioned high in the center of the chamber, above the test specimens.
Multiple spray nozzles spray the salt solution mist directly and evenly
onto the test specimens below.
Spray bar nozzles equipped with integrated filters rotate within the
chamber at a custom speed back and forth,
to optimize the precipitation for specific test protocols.
Ultra-fast heating system with ramp function
Temperature range 35°C to 70°C + humidity
Enables ultra-fast transition time from 35°C salt spray mist to 70°C
drying within 15 minutes.
Also allows for maximum temperature and relative humidity of
80°C and 80% relative humidity to be reached.
35 °C salt spray mist (100% relative humidity) to 70 °C air drying < 30 min.
70 °C air drying to 50 °C > 95% relative humidity < 20 min.
50 °C > 95% relative humidity to salt spray mist 35 °C
(100% relative humidity) < 30 min.
60 °C air drying to -20 °C < 60 min.
Innovative design, the oscillating spray bar and air supply
The system can be quickly and easily removed to allow the chamber to comply with other
cyclical corrosion test standards.
Specifications
Atmosfär Test Chambers 1300 liters
Chamber capacity: 1300 liters
Chamber weight: 340 kg
Load: 150 kg
Mounting format: Floor-standing
Load threshold 800 mm
Chamber outer dimensions, max. width 2025 mm depth 1316 mm height 1965 mm
Chamber inner dimensions, max. width 1300 mm depth 980 mm height 1500 mm
Cooling unit:
Separate unit, mounted on the left side of the Atmosfär chamber.
Width 650 mm depth 870 mm height 1100 mm
Salt solution tank: 160 liters/
One unit is supplied to deliver saltwater to the spray bar.
Width 840 mm depth 500 mm height 790 mm
Chamber construction:
Fiberglass reinforced plastic, polypropylene, and ABS parts.
Elix Essential 3 Water Treatment System: ZLXE0030WW
Product information:
Daily water consumption of the products 60 l/day
Feed water type Drinkable tap water
Water quality Type 2
Application:
A pure water solution with integrated Elix technology for your basic requirements.
Physicochemical information
Conductivity (µS/cm) < 0.2 µS/cm
Flow rate (l/h) 3 l/h
Flow rate 3 l/h
Product water resistance @ 25 °C Typically 10–15 MΩ·cm
TOC of product water < 30 ppb
Voltage: 100–230 V/50–60 Hz
Dimensions: Height 47 cm Width 26.8 cm Depth 33.9 cm
Operating weight: 17.2 - 18.5 kg
Electrical connection values: AT1300IP/3
Voltage 380-440V
Frequency 50 Hz 20 A 7189 W
3 phases 380-440 V
Weight: 340 kg
Condition: used
Scope of delivery: (See image)
(Changes and errors in technical data are reserved!)
For further questions, we would be happy to answer them by phone.

Technical characteristic Value
Item ID 24500
Condition Used
Model Ascott AT1300IP/3
Manufacturer Ascott
Weight 400000 g
document.querySelector("form > input[data-mail='subject']").value = "Webshop - Frage zu Artikel - Artikel-ID: " + vueApp.$store.state.items[24500].variation.documents[0].data.item.id;

EU Responsible Person

Weiss Technik GmbH
Greizer Straße 41-49
35447 Reiskirchen Germany
info@weiss-technik.com
+49 6408 84-0

Manufacturer details

Ascott
Ascott Analytical Equipment Limited
United Kingdom

Switch to environmentally friendly refrigerant

The following work can be performed on climate chambers. Please contact us if you are interested.

 

  • Extraction and proper disposal of the previous refrigerant
  • Replacement of necessary components, if unavoidable
  • Evacuate the circuit and refill it with R 452A, for example.
  • Checking / adjusting overheating in expansion valves
  • Leak test
  • Test run of the system

Salt spray chambers